Instruction Manual HI 9524 Portable Waterproof pH Meter http://www.hannainst.
Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction manual carefully before using the instrument. It will provide you with the necessary information for a correct use of the instrument, as well as a precise idea of its versatility. directives. These instruments are in compliance with the TABLE OF CONTENTS PRELIMINARY EXAMINATION ........................................................ 3 GENERAL DESCRIPTION ...............................................................
PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, notify your Dealer immediately. Note: Save all packing materials until you are sure that the instrument functions correctly. Any damaged or defective item must be returned in its original packing materials together with the supplied accessories.
FUNCTIONAL DESCRIPTION 1) 2) 3) 4) 5) 6) 7) 8) 9) 10) 11) 12) BNC electrode connector Temperature probe socket Liquid Cristal Display ON/OFF key to turn the meter on or off CAL key to enter or exit calibration mode CFM key to confirm calibration ϰC and аC keys for manual temperature setting, or selecting pH buffer value MEM key to store pH value in memory MR key to recall the stored value from memory User-friendly graphic symbols Primary display Secondary display 4
SPECIFICATIONS Range 0.00 to 14.00 pH; 0.0 to 100.0°C Resolution 0.01 pH; 0.1°C Accuracy (@20°C/68°F) ±0.01 pH; ±0.5°C Typical EMC Deviation ±0.01 pH; ±0.1°C Calibration Automatic with 5 memorized standard buffers (4.01, 6.86, 7.01, 9.18, 10.
OPERATIONAL GUIDE INITIAL PREPARATION The meter is supplied complete with batteries. Remove the back cover, unwrap the batteries and install them while paying attention to polarity. To prepare the instrument for use connect the pH electrode and the temperature probe to the BNC and temperature sockets on the top of the instrument. The temperature probe can be used independently to take temperature measurements, or it can be used in conjunction with the pH electrode to utilize the meter's ATC capability.
meter to measure the pH accurately, temperature must be taken in consideration. If the sample temperature is quite different from the temperature at which the pH electrode was kept, allow a few minutes for a perfect thermal equilibrium between them. To use the meter's Automatic Temperature Compensation feature, submerge the temperature probe into the sample as close to the electrode as possible and wait for a couple of minutes.
p H CALIBRATION It is recommended to calibrate the instrument frequently, especially if high accuracy is required. The instrument should be recalibrated for pH: a) Whenever the pH electrode or temperature probe is replaced. b) At least once a month. c) After testing aggressive chemicals. d) If greatest accuracy is required. e) Whenever the batteries have been replaced. PREPARATION 04 70 HI Pour small quantities of pH 7.01 and pH 4.01 solution into two clean beakers.
the Standard 7.01 or the NIST 6.86 are the most appropriate for this purpose even though both meters can be calibrated with any of the 5 memorized calibration solutions. • Immerse the pH electrode into a pH buffer solution (e.g. pH 6.86) and stir gently. Note: the electrode should be submerged approximately 4 cm (1½") into the solution. With the temperature probe located as close to it as possible. • Press the CAL key. The "CAL" and " 1" indicators will be displayed.
• Press the CAL key and the calibration process is ended with only the offset of the meter calibrated. For best accuracy however, do not press CAL and proceed with a two-point calibration. TWO-POINT CALIBRATION After calibrating at 6.86 or 7.01 pH (see above), for the second buffer use pH 4.01 if you are going to measure acidic samples, and pH 9.18 or pH 10.01 if you are going to measure alkaline solutions.
p H VALUES AT VARIOUS TEMPERATURES Temperature has an effect on pH. The calibration buffer solutions are effected by temperature changes to a lesser degree than normal solutions. For manual temperature calibration please refer to the following chart. For instance, if the buffer temperature is 25°C, the display should show pH 4.01, 7.01 or 10.01 at pH 4, 7 or 10 buffers, respectively. At 20°C, the display should show pH 4.00, 7.03 or 10.06. The meter reading at 50°C will then be 4.06, 6.98 or 9.82.
TROUBLESHOOTING GUIDE Symptoms Problem Solution The meter is slow in re- The electrode is not work- Leave the electrode in a storsponding or gives faulty ing or the reference age solution after cleaning readings the junction. If problem perjunction is clogged sists, replace the electrode The meter does not ac- Out of order pH electrode Follow the cleaning procecept the 2 nd buffer dure.
ELECTRODE CONDITIONING & MAINTENANCE * Not present in gel electrodes. PREPARATION Remove the protective cap. DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is normal with electrodes and they will disappear when rinsed with water. During transport tiny bubbles of air may form inside the glass bulb affecting proper functioning of the electrode. These bubbles can be removed by "shaking down" the electrode as you would do with a glass thermometer.
For refillable electrodes: If the filling solution (electrolyte) is more than 2½ cm (1") below the fill hole, add HI 7082 or HI 8082 3.5M KCl Electrolyte Solution for double junction or HI 7071 or HI 8071 3.5M KCl+AgCl Electrolyte Solution for single junction electrodes. For a faster response, unscrew the fill hole screw during measurements. For AmpHel®electrodes: If the electrode does not respond to pH changes, the battery is run down and the electrode should be replaced.
CLEANING PROCEDURE General Soak in Hanna HI 7061 or HI 8061 General Cleaning Solution for approximately ½ hour. Removal of films, dirt or deposits on the membrane/junction: - Protein Soak in Hanna HI 7073 or HI 8073 Protein Cleaning Solution for 15 minutes. - Inorganic Soak in Hanna HI 7074 or HI 8074 Inorganic Cleaning Solution for 15 minutes. - Oil/grease Rinse with Hanna HI 7077 or HI 8077 Oil and Fat Cleaning Solution.
BATTERY REPLACEMENT & AUTO-OFF If the batteries become weak the meter will display "LOW BAT". The warning message is to alert the user that the display will be shut-off after about 5 hours of use. After that, the meter shuts down the display, preventing the taking of erroneous measurements due to low voltage. Battery replacement must only take place in a non hazardous area using the battery types specified in this instruction manual.
ACCESSORIES pH CALIBRATION SOLUTIONS HI 70004P HI 70007P HI 70010P HI 7004L HI 7006L HI 7007L HI 7009L HI 7010L pH 4.01 Buffer Sachets, 20mL, 25 pcs pH 7.01 Buffer Sachets, 20mL, 25 pcs pH 10.01 Buffer Sachets, 20mL, 25 pcs pH 4.01 Buffer Solution, 460 mL pH 6.86 Buffer Solution, 460 mL pH 7.01 Buffer Solution, 460 mL pH 9.18 Buffer Solution, 460 mL pH 10.01 Buffer Sol., 460 mL pH CALIBRATION SOLUTIONS IN FDA APPROVED BOTTLES HI 8004L HI 8006L HI 8007L HI 8009L HI 8010L pH 4.
REFILL ELECTROLYTE SOLUTIONS IN FDA APPROVED BOTTLES HI 8071 HI 8072 HI 8082 3.5M KCl + AgCl Electrolyte, 4x50mL, for single junction electrodes 1M KNO3 Electrolyte, 4x50 mL 3.5M KCl Electrolyte, 4x50 mL, for double junction electrodes pH ELECTRODES All electrodes part numbers ending in B are supplied with a BNC connector and 1 m (3.3') cable. HI 1043B Glass-body, double junction, refillable, combination pH electrode. Use: strong acid/alkali.
HI 1413B Glass-body, single junction, flat tip, Viscolene, nonrefillable, combination pH electrode. Use: surface measurement. OTHER ACCESSORIES Electrode holder Calibration screwdriver (20 pcs) Temperature probe with 1 m (3.3') screened cable pH and ORP electrode simulator with 1 m (3.3') coaxial cable ending in female BNC connectors HI 931001 pH and ORP electrode simulator with LCD and 1 m (3.
ELECTRODE APPLICATION REFERENCE GUIDE Application 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. 21. 22. 23. 24. 25. 26. 27. 28. 29. 30. 31. 32. 33. 34. 35. 36. 37. 38. 39. 40. 41.
WARRANTY All Hanna Instruments meters are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The electrodes and the probes are guaranteed for a period of six months. This warranty is limited to repair or replacement free of charge. Damage due to accident, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument.
CE DECLARATION OF CONFORMITY Recommendations for Users Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential area could cause unacceptable interferences to radio and TV equipment, requiring the operator to take all necessary steps to correct interferences. The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid touching this glass bulb at all times.
HANNA LITERATURE Hanna publishes a wide range of catalogs and handbooks for an equally wide range of applications. The reference literature currently covers areas such as: • Water Treatment • Process • Swimming Pools • Agriculture • Food • Laboratory • Thermometry and many others. New reference material is constantly being added to the library. For these and other catalogs, handbooks and leaflets, contact your dealer or the Hanna Customer Service nearest to you.
MAN9524 10/01 h t t p : / / w w w . h a n n a i n s t .