The HI 93724 meter measures cyanuric acid and pH content in water
and wastewater in the following ranges: 0 to 80 mg/L for Cyanuric
Acid (CYS) and 5.9 to 8.5 for pH.
SPECIFICATIONS
Range CYS 0 to 80 mg/L
pH 5.9 to 8.5
Resolution CYS 1 mg/L
pH 0.1
Accuracy CYS ±1 mg/L ±15% of reading
pH ±0.1
Typical EMC CYS ±1 mg/L
Deviation pH ±0.1
Light Source Light Emitting Diode @ 555 nm
Method For pH, Phenol Red method. The reaction with
reagent causes a red tint in the sample.
For cyanuric acid, adaptation of the turbidimet-
ric method. The reaction between cyanuric acid
and the reagent causes a white suspension in
the sample.
REQUIRED REAGENTS
Code Unit Description Quantity
HI 93710-0 pH Phenol red 5 drops
HI 93722-0 CYS Powder Reagent 1 packet
44
• Fill one graduated mixing cylinder up to the
25 mL mark with deionized water.
• Add the content of one packet of HI 93746 TPTZ reagent, close
the cylinder and agitate for 30 seconds. This is the blank.
• Fill a cuvet with 10 mL of the blank up
to the mark and let it sit for 3 minutes.
• Replace the cap, insert the blank into
the holder and make sure that the notch
on the cap is positioned securely into
the groove.
• Press ZERO and "SIP" will appear on the display.
• Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
• Fill one graduated mixing cylinder up to the
25 mL mark with the sample.
• Add the content of one packet of HI 93746 TPTZ reagent, close
the cylinder and agitate for 30 seconds.