The HI 93710 meter measures both pH and chlorine (Cl
2
, free and
total) content in water and wastewater in the following ranges:
pH 5.9 to 8.5 pH units
Free Chlorine 0.00 to 2.50 mg/L (ppm)
Total Chlorine 0.00 to 3.50 mg/L (ppm).
SPECIFICATIONS
Range pH 5.9 to 8.5
Free Cl
2
0.00 to 2.50 mg/L
Total Cl
2
0.00 to 3.50 mg/L
Resolution 0.1 pH/0.01 mg/L Cl
2
Accuracy ±0.1 pH
±0.03 mg/L ±3% of Cl
2
reading
Typical EMC ±0.2 pH
Deviation ±0.02 mg/L Cl
2
Light Source Light Emitting Diode @ 555 nm
Method Adaptation of the EPA recommended DPD method
330.5 for chlorine analysis. The reaction with
reagents causes a pink tint in the sample.
For pH, Phenol red method. The reaction with
reagents causes a red tint in the sample.
REQUIRED REAGENTS
Code Unit Description Quantity
HI 93710-0 pH Phenol red 5 drops
HI 93701-0 Free Cl
2
DPD 1 packet
HI 93710
pH/Chlorine
46
pp
pp
p
H MEASUREMENTH MEASUREMENT
H MEASUREMENTH MEASUREMENT
H MEASUREMENT
Fill the cuvet up to the mark with 10 mL of
unreacted sample and replace the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP” will appear on the display.
Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed and
ready for measurement.
Remove the cuvet and add 5
drops of the HI 93710 Phenol
Red Indicator. Replace the cap
and swirl the solution.
Reinsert the cuvet into the instrument.
Press the READ key and “SIP” will appear on the display during
measurement.
The instrument directly displays the pH measured value on the
Liquid Crystal Display.
ZERO
pH
10 mL
x 5