User Guide

18
Press READ. "SIP" will appear during measurement.
The instrument directly displays concentration in µg/L of manga-
nese on the display.
Note: a temperature above 30°C may cause turbidity. In this case,
before zeroing and taking readings, add 2-3 drops of Dispers-
ing Agent (HI 93703-51) to each cuvet and swirl until
complete dissolution of the turbidity.
INTERFERENCES
Manganese
Interference may be caused by:
Aluminum above 20 mg/L
Cadmium above 10 mg/L
Calcium above 200 mg/L as CaCO
3
Cobalt above 20 mg/L
Copper above 50 mg/L
Iron above 10 mg/L
Lead above 0.5 mg/L
Magnesium above 100 mg/L as CaCO
3
Nickel above 40 mg/L
Zinc above 15 mg/L.
HI 93742
Iron LR/Manganese LR
35
Reinsert the cuvet into the instrument.
Wait for approximately 30 seconds and then press READ. The
display will show “SIP” during measurement.
The instrument directly displays concentration in mg/L of free
chlorine on the Liquid Crystal Display.
Note: Free and total chlorine have to be measured separately with
fresh unreacted samples following the above procedure if both
values are requested.
INTERFERENCES
Interference may be caused by:
Bromine
Iodine
Fluorine
Ozone
Oxidized manganese and Chromium
In case of water with hardness greater than 500 mg/L CaCO
3
, shake
the sample for approximately 1 minute after adding the reagent.
In case of water with alkalinity greater than 250 mg/L CaCO
3
or
acidity greater than 150 mg/L CaCO
3
, the color of the sample could
disappear or develop only partially. To resolve this, neutralize the
sample with diluted HCl or NaOH.
Free Chlorine