User Guide
18
• Press READ. "SIP" will appear during measurement.
• The instrument directly displays concentration in µg/L of manga-
nese on the display.
Note: a temperature above 30°C may cause turbidity. In this case,
before zeroing and taking readings, add 2-3 drops of Dispers-
ing Agent (HI 93703-51) to each cuvet and swirl until
complete dissolution of the turbidity.
INTERFERENCES
Manganese
Interference may be caused by:
Aluminum above 20 mg/L
Cadmium above 10 mg/L
Calcium above 200 mg/L as CaCO
3
Cobalt above 20 mg/L
Copper above 50 mg/L
Iron above 10 mg/L
Lead above 0.5 mg/L
Magnesium above 100 mg/L as CaCO
3
Nickel above 40 mg/L
Zinc above 15 mg/L.
HI 93742
Iron LR/Manganese LR
35
• Reinsert the cuvet into the instrument.
• Wait for approximately 30 seconds and then press READ. The
display will show “SIP” during measurement.
• The instrument directly displays concentration in mg/L of free
chlorine on the Liquid Crystal Display.
Note: Free and total chlorine have to be measured separately with
fresh unreacted samples following the above procedure if both
values are requested.
INTERFERENCES
Interference may be caused by:
Bromine
Iodine
Fluorine
Ozone
Oxidized manganese and Chromium
In case of water with hardness greater than 500 mg/L CaCO
3
, shake
the sample for approximately 1 minute after adding the reagent.
In case of water with alkalinity greater than 250 mg/L CaCO
3
or
acidity greater than 150 mg/L CaCO
3
, the color of the sample could
disappear or develop only partially. To resolve this, neutralize the
sample with diluted HCl or NaOH.
Free Chlorine