Add the content of one packet of
HI 93748A ascorbic acid to each
cuvet, replace the caps and shake
gently.
Add 0.2 mL of the
HI 93748B alkaline-
cyanide reagent
solution to each cu-
vet, replace the
caps and shake gen-
tly.
Add 1 mL of the HI 93748C
0.1% PAN indicator solution
to each cuvet, replace the
caps and shake gently.
Place the cuvet with the reacted deion-
ized water (blank) into the holder
and ensure that the notch on the cap
is positioned securely into the groove.
Wait for 2 minutes and press ZERO. "SIP" will appear during
zeroing.
Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
Insert the second cuvet with the reacted
sample into the instrument.
ZERO
12
3
6
9
2’
HI 93742
Iron LR/Manganese LR36
TOTAL CHLORINE MEASUREMENTTOTAL CHLORINE MEASUREMENT
TOTAL CHLORINE MEASUREMENTTOTAL CHLORINE MEASUREMENT
TOTAL CHLORINE MEASUREMENT
H
I 9
3
7
1
1
Fill the cuvet up to the mark with 10 mL of
unreacted sample and replace the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and "SIP" will appear on the display.
Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
Powder reagents procedure
Remove the cuvet and add one packet of DPD Total
Chlorine reagent. Replace the cap and shake gently.
Reinsert the cuvet into the instrument.
Wait for 2 minutes and 30 seconds and then press READ. The
display will show "SIP" during measurement.
The instrument directly displays concentration in mg/L of total
chlorine on the Liquid Crystal Display.
ZERO
Total Chlorine
10 mL