The HI 93742 meter measures the iron and manganese content in
water, wastewater and seawater (iron only) in the following ranges:
Iron 0 to 400 µg/L (ppb)
Manganese 0 to 300 µg/L (ppb).
SPECIFICATIONS
Range Fe 0 to 400 µg/L
Mn 0 to 300 µg/L
Resolution 1 µg/L
Accuracy Fe ±10 µg/L ±8% of reading
Mn ±2 µg/L ±3% of reading
Typical EMC ±1 µg/L
Deviation
Light Source Light Emitting Diode @ 555 nm
Method Adaptation of the TPTZ method for iron and
PAN method for manganese. The reaction be-
tween iron or manganese and the reagents
respectively causes a blue or an orange tint in
the sample.
REQUIRED REAGENT
Code Description Quantity
Iron
HI 93746-0 TPTZ reagent 2 packets
HI 93742
Iron LR/Manganese LR
HI 93742 - HI 93742 -
HI 93742 - HI 93742 -
HI 93742 -
Iron Low Range andIron Low Range and
Iron Low Range andIron Low Range and
Iron Low Range and
Manganese Low RangeManganese Low Range
Manganese Low RangeManganese Low Range
Manganese Low Range
38
Fill the cuvet up to the mark with 10 mL of
sample and replace the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and "SIP" will appear on the display.
Wait for a few seconds and the display will
show "0". Now the meter is zeroed and ready
for measurement.
Fill a beaker up to the
25 mL mark with the
sample, add the con-
tent of one packet of
HI 93722 reagent and
stir gently to mix.
Fill a second cuvet with 10 mL of the reacted
sample up to the mark. Replace the cap.
Reinsert the cuvet into the instrument.
Wait for 45 seconds and press READ. "SIP" will appear during
measurement.
The instrument directly displays concentration in mg/L of cyanuric
acid on the Liquid Crystal Display.