User Guide

11
HI 93725
pH & Total Hardness
HI 93725 - pH & Total HardnessHI 93725 - pH & Total Hardness
HI 93725 - pH & Total HardnessHI 93725 - pH & Total Hardness
HI 93725 - pH & Total Hardness
The HI 93725 meter measures pH and Magnesium/Calcium and Total
Hardness content in water and wastewater.
SPECIFICATIONS
Range pH 5.9 to 8.5
Mg Hardness 0.00 to 2.00 mg/L
Ca Hardness 0.00 to 2.70 mg/L
Total Hardness 0.00 to 4.70 mg/L
Resolution pH 0.1
Hardness 0.01 mg/L
Accuracy pH ±0.1
Hardness ±0.11 mg/L ±5% of reading)
Typical EMC pH ±0.1
Deviation Hardness ±0.02 mg/L
Light Source Light Emitting Diode @ 555 nm
Method For pH, Phenol red method. The reaction with
reagent causes a red tint in the sample.
For total hardness, adaptation of the Standard
Methods for the Examination of Water and
Wastewater, 18
th
edition, Calmagite/colorimet-
ric method. The reaction between hardness and
reagents causes a maroon tint in the sample.
42
IODINE MEASUREMENTIODINE MEASUREMENT
IODINE MEASUREMENTIODINE MEASUREMENT
IODINE MEASUREMENT
ZERO
Fill the cuvet up to the mark with 10 mL of
unreacted sample and replace the cap.
Place the cuvet into the holder and en-
sure that the notch on the cap is
positioned securely into the groove.
Press ZERO and "SIP" will appear on the display.
Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
Remove the cap and add the content of
one packet of HI 93718 reagent. Re-
place the cap and shake gently.
Reinsert the cuvet into the instrument.
Wait for 2 minutes and 30 seconds and press READ. "SIP" will
appear during measurement.
The instrument directly displays concentration in mg/L of iodine
on the Liquid Crystal Display.
Iodine
10 mL