User Guide
11
HI 93725
pH & Total Hardness
HI 93725 - pH & Total HardnessHI 93725 - pH & Total Hardness
HI 93725 - pH & Total HardnessHI 93725 - pH & Total Hardness
HI 93725 - pH & Total Hardness
The HI 93725 meter measures pH and Magnesium/Calcium and Total
Hardness content in water and wastewater.
SPECIFICATIONS
Range pH 5.9 to 8.5
Mg Hardness 0.00 to 2.00 mg/L
Ca Hardness 0.00 to 2.70 mg/L
Total Hardness 0.00 to 4.70 mg/L
Resolution pH 0.1
Hardness 0.01 mg/L
Accuracy pH ±0.1
Hardness ±0.11 mg/L ±5% of reading)
Typical EMC pH ±0.1
Deviation Hardness ±0.02 mg/L
Light Source Light Emitting Diode @ 555 nm
Method For pH, Phenol red method. The reaction with
reagent causes a red tint in the sample.
For total hardness, adaptation of the Standard
Methods for the Examination of Water and
Wastewater, 18
th
edition, Calmagite/colorimet-
ric method. The reaction between hardness and
reagents causes a maroon tint in the sample.
42
IODINE MEASUREMENTIODINE MEASUREMENT
IODINE MEASUREMENTIODINE MEASUREMENT
IODINE MEASUREMENT
ZERO
• Fill the cuvet up to the mark with 10 mL of
unreacted sample and replace the cap.
• Place the cuvet into the holder and en-
sure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and "SIP" will appear on the display.
• Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
• Remove the cap and add the content of
one packet of HI 93718 reagent. Re-
place the cap and shake gently.
• Reinsert the cuvet into the instrument.
• Wait for 2 minutes and 30 seconds and press READ. "SIP" will
appear during measurement.
• The instrument directly displays concentration in mg/L of iodine
on the Liquid Crystal Display.
Iodine
10 mL