User Guide
10
REAGENT SETS
HI 93710-01 Reagents for 100 pH tests
HI 93710-03 Reagents for 300 pH tests
HI 93722-01 Reagents for 100 CYS tests
HI 93722-03 Reagents for 300 CYS tests
For other accessories see page 48.
MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/OFF.
• The meter will automatically default to pH mea-
surement mode.
• When the LCD displays “- - -”, it is ready.
pH MEASUREMENTS
In order to perform pH measurements, follow the procedure on
page 46.
CYANURIC ACID MEASUREMENTS
• Press the pH/CYS key to select the cyanuric acid scale. “pH” will
disappear from the LCD.
• Follow the procedure on page 38.
ON
OFF
pH
HI 93724
Cyanuric A. & pH
43
INTERFERENCES
Interference may be caused by:
Bromine
Chlorine
Ozone
Oxidized forms of chromium and manganese
Alkalinity above 300 mg/L and acidity above 150 mg/L as CaCO
3
require neutralization since the color could instantly fade. To resolve
this, neutralize the sample with diluted HCl or NaOH.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake
the sample for approximately 1 minute after adding the reagent.
Iodine