User Guide

11
HOW TO ENSURE ACCURATE MEASURE-
MENTS
Each time the cuvet is used, tighten the cap
to the same degree.
Discard the sample soon after the reading is
taken to avoid permanently clouding the glass.
All glassware used to contain the standards
and the samples should be maintained clean,
washed with HI93703-50 cleaning solution
and rinsed with HI 93703-0 or turbidity-free
water.
Collect the samples in clean glass or plastic
bottles, fit stoppers and perform the analysis
quickly. If unavoidable, store the sample in
a cool, dark place, but not for longer than 24
hours (the sample needs to be kept at room
temperature prior to the analysis).
To obtain a representative sample, gently,
but thoroughly, mix it before samples are
taken. Do not shake (to prevent air bubbles)
and do not let the sample settle.
It is recommended to monthly calibrate the
meter with the supplied HI93703-10 @10
FTU* standard or more frequently for great-
est accuracy.
Before inserting vials
into the instrument,
wipe them with
HI93703-70 soft, lint-
free tissue. Handle vi-
als so that no finger-
prints can get on the
areas where light
passes (approximately
2 cm/1 inch from the
bottom of the vial).
If you experience any problems in taking mea-
surements, please contact your dealer or the
nearest Hanna Instruments Customer Service
Department.
* 1 FTU = 1 NTU