User Guide

15
* 1 FTU = 1 NTU
hours (the sample needs to be kept at room
temperature prior to the analysis).
To obtain a representative sample, gently,
but thoroughly, mix it before samples are
taken. Do not shake (to prevent air bubbles)
and do not let the sample settle.
It is recommended to calibrate the meter with
the supplied HI93703-10 @10 FTU* stan-
dard at least once a month or more frequently
for greater accuracy.
Before inserting vials into the instrument,
wipe them with HI 93703-70 or a soft, lint-
free tissue. Handle vials so that no finger-
prints get on the areas where light passes
(approx. the bottom 2 cm/1 inch).
If you experience any problems in taking mea-
surements, contact your dealer or the nearest
Hanna Customer Service Center.
SOURCES OF INTERFERENCE
Presence of floating debris and coarse sedi-
ments which settle out rapidly will give false
readings.
The infrared light source used for HI 93703-
11 turbidity meter, complies with the ISO
7027 International Standard and can effec-
tively minimize any errors due to colored
dissolved substances. This effect, called
"true color", is a common interference for
most commercially available instruments
operating in the range of visible light.
Air bubbles and the effect of vibrations that
disturb the surface of the sample may give
false results and should be avoided.
Dirty glassware could also affect readings
along with scratched or dented vials.
16
To check the date of last calibration, simply
press the GLP/CAL key. Press again to
toggle between date and time.
To make sure that the meter is calibrated, take
a measure of a standard solution.
The instrument can be calibrated at two or three
points and a monthly calibration is recom-
mended.
CALIBRATION PROCEDURE
Turn the meter on and
wait for the display to
show "----".
Press the ALT & CAL
buttons together. The
"CAL" message will
blink on the display 3
times. The meter then
enters the calibration
mode, displaying "0.00
cl" and prompting the
user to insert the 0.00
FTU standard.
Place the 0.00 FTU stan-
dard in the cuvet holder
Press CAL, SIP and
CL will start blinking
LOGGING
MICROPROCESSOR
TURBIDITY METER
LOGGING
MICROPROCESSOR
TURBIDITY
METER
HI 93703-11
VIEW
STO
CLR
GLP
CAL
DATE
READ
ALT
OFF
ON
CALIBRATION
0.00
cl
sip
cl