User Guide
13
• Press right arrow to scroll
to month and day, use
UP/DOWN to set.
• Press right arrow again
to scroll to time, use UP/
DOWN to set.
• Press right arrow again
to save and exit date and
time setup.
NOTE: Press ALT & DATE to leave this mode
without saving.
ERROR CODES:
Every time the meter is switched on, the Real
Time Clock and EEPROM are tested, and if an
error is found the corresponding error code will
be displayed.
The list of error codes is as follows:
• NO cover error
(check cuvet position)
• Calibration error
(check calibration
standard value)
LOGGING
MICROPROCESSOR
TURBIDITY METER
LOGGING
MICROPROCESSOR
TURBIDITY
METER
HI 93703-11
VIEW
STO
CLR
GLP
CAL
DATE
READ
ALT
OFF
ON
cap
err1
LOGGING
MICROPROCESSOR
TURBIDITY METER
LOGGING
MICROPROCESSOR
TURBIDITY
METER
HI 93703-11
VIEW
STO
CLR
GLP
CAL
DATE
READ
ALT
OFF
ON
14
• Real Time Clock
error ∗
• EEPROM error∗
• Internal communica-
tion error∗
• Internal Bus error∗
∗
Contact the HANNA office nearest to you.
err2
err3
err4
err5
ENSURE ACCURATE MEASUREMENTS
• Each time the cuvet is used, tighten the cap
to the same degree.
• Discard the sample soon after the reading is
taken to avoid permanently clouding the glass.
• All glassware used to contain the standards
and the samples should be maintained clean,
washed with the HI93703-50 cleaning solution
and rinsed with HI 93703-0 or turbidity-free
water.
• Collect the samples in clean glass or plastic
bottles, fit stoppers and perform the analysis
quickly. If necessary, store the sample in a
cool, dark place, for up to a maximum of 24