Owner's manual

98
HI 93532 HI 93532N
Keyboard Functions:
ON/OFF : turn the meter ON and OFF
T1/T2 : select the reading channel (T1, T2 or T1-T2)
HOLD : freeze the reading on display
CLR : clear the HI and LO values
BL (HI 93532N only) : toggle the backlight ON and OFF.
CAL (HI 93532N only) : press and hold for about 5 seconds to enter
the Calibration mode (with reading within ±3°C); press to abort
the Calibration mode
SPECIFICATIONS OF
HI 93532 & HI 93532N
Main features:
Two independent measurement channels, T1 and T2
Use HI 766 K-thermocouple probes
Display of temperature in degrees Celsius or Fahrenheit
Extended range -200 to 1371°C or -328 to 2500°F
HOLD function
HI & LO (Max & Min) temperature values displayed on the LCD
Remaining battery life indication / low battery detection
Waterproof casing
Auto-off capability
Backlight feature (HI 93532N only)
User calibration at 0°C (HI 93532N only)
SPECIFICATIONS
Range (*) -200.0 to 999.9°C / 1000 to 1371°C
-328.0 to 999.9°F / 1000 to 2500°F
Resolution 0.1°C (-149.9 to 999.9°C)
0.2°C (-200.0 to -150.0°C) / 1°C (outside)
0.1°F (-24.9 to 999.9°F)
0.2°F (-249.9 to -25.0°F)
0.3°F (-328.0 to -250.0°F) / 1°F (outside)
Accuracy ±0.5°C (-100.0 to 999.9°C) / ±1°C (outside)
(@20°C/68°F) ±1°F (-148.0 to 999.9°F) / ±1.5°F (outside)
excluding probe error
Typical EMC ±3°C / ±6°F
Deviation with HI 766 K-thermocouple probe
Battery 3x1.5V AA (IEC LR6) batteries,
approx. 500 hours of continuous use (BL off)
Auto-off user selectable: 60 min or disabled
Environment -10 to 60°C (14 to 140°F); RH 100%
Dimensions 150 x 80 x 36 mm (5.9 x 3.1 x 1.4")
Weight 235 g (8.3 oz.)
(*) Range may be limited by probe.