Instruction Manual HI 93530 • HI 93530N HI 93531 • HI 93531N HI 93531R HI 93532 • HI 93532N HI 93532R K-Thermometers w w w . h a n n a i n s t .
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for correct use of the instrument, as well as a more precise idea of its versatility. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. directives.
SPECIFICATIONS OF HI 93530 & HI 93530N HI 93530 SPECIFICATIONS Range (*) HI 93530N -200.0 to 999.9°C / 1000 to 1371°C -328.0 to 999.9°F / 1000 to 2500°F Resolution 0.1°C (-149.9 to 999.9°C) 0.2°C (-200.0 to -150.0°C) / 1°C (outside) 0.1°F (-24.9 to 999.9°F) 0.2°F (-249.9 to -25.0°F) 0.3°F (-328.0 to -250.0°F) / 1°F (outside) Accuracy ±0.5°C (-100.0 to 999.9°C) / ±1°C (outside) (@20°C/68°F) ±1°F (-148.0 to 999.9°F) / ±1.
SPECIFICATIONS OF HI 93531 & HI 93531N HI 93531 SPECIFICATIONS Range (*) HI 93531N -200.0 to 999.9°C / 1000 to 1371°C -328.0 to 999.9°F / 1000 to 2500°F Resolution 0.1°C (-149.9 to 999.9°C) 0.2°C (-200.0 to -150.0°C) / 1°C (outside) 0.1°F (-24.9 to 999.9°F) 0.2°F (-249.9 to -25.0°F) 0.3°F (-328.0 to -250.0°F) / 1°F (outside) Accuracy ±0.5°C (-100.0 to 999.9°C) / ±1°C (outside) (@20°C/68°F) ±1°F (-148.0 to 999.9°F) / ±1.
SPECIFICATIONS OF HI 93532 & HI 93532N HI 93532 SPECIFICATIONS Range (*) HI 93532N -200.0 to 999.9°C / 1000 to 1371°C -328.0 to 999.9°F / 1000 to 2500°F Resolution 0.1°C (-149.9 to 999.9°C) 0.2°C (-200.0 to -150.0°C) / 1°C (outside) 0.1°F (-24.9 to 999.9°F) 0.2°F (-249.9 to -25.0°F) 0.3°F (-328.0 to -250.0°F) / 1°F (outside) Accuracy ±0.5°C (-100.0 to 999.9°C) / ±1°C (outside) (@20°C/68°F) ±1°F (-148.0 to 999.9°F) / ±1.
SPECIFICATIONS OF HI 93531R & HI 93532R HI 93531R HI 93532R SPECIFICATIONS Range (*) -200.0 to 999.9°C / 1000 to 1371°C -328.0 to 999.9°F / 1000 to 2500°F Resolution 0.1°C (-149.9 to 999.9°C) 0.2°C (-200.0 to -150.0°C) / 1°C (outside) 0.1°F (-24.9 to 999.9°F) 0.2°F (-249.9 to -25.0°F) 0.3°F (-328.0 to -250.0°F) / 1°F (outside) Accuracy ±0.5°C (-100.0 to 999.9°C) / ±1°C (outside) (@20°C/68°F) ±1°F (-148.0 to 999.9°F) / ±1.
OPERATIONAL GUIDE Note: When the reading goes over-range or the probe is removed, the Hi and Lo values display dashes until cleared. To switch the instrument ON, press the ON/OFF key. The thermometer will carry out a self diagnostic test routine, the LCD will show all segments for a few seconds (or as long as ON/OFF is held), followed by the percentage indication of the remaining battery life. CLEAR FUNCTION (not for HI 93530 and HI 93530N) The thermometer then enters normal measurement mode.
BACKLIGHT FEATURE (not for HI 93530, HI 93531 and HI 93532) The backlight feature can be easily activated through the keyboard by pressing the BL (or ALT+BL) key. Note: The backlight automatically shuts off after approximately 1 minute with no buttons pressed. BATTERY COMPARTMENT See "Battery replacement" section for back cover removal/installation.
The average value is lost and will not restore even if the condition restores. To start the average cycle again, either press CLR or reenter the average mode. • While in T1-T2 (HI 93532R only), pressing DISP will switch between 4 displays: Note: Pressing T1/T2 (HI 93532R only) to change the channel will not alter the information chosen to be displayed for each channel (DISP). For example, if T1 is in average mode, the average will continue to be calculated even if the user press T1/T2 to see T2.
USER CALIBRATION (Not for HI 93530, HI 93531 and HI 93532 models) The meters can be calibrated at 0°C by using an ice bath. • Prepare an ice bath with approximately equal volumes of distilled water and chopped ice made from distilled water. • Immerse the temperature probe in the center of the ice bath, taking care not to touch the ice with the probe tip. • Ensure that the meter is measuring a temperature within ±3°C. • To enter the Calibration mode, press and hold the CAL button for about 5 seconds.
SALES AND TECHNICAL SERVICE CONTACTS Australia Tel. (03) 9769.0666 • Fax (03) 9769.0699 China Tel. (10) 88570068 • Fax (10) 88570060 Egypt Tel. & Fax (02) 2758.683 Germany Tel. (07851) 9129-0 • Fax (07851) 9129-99 Greece Tel. (210) 823.5192 • Fax (210) 884.0210 Indonesia Tel. (21) 4584.2941 • Fax (21) 4584.2942 Japan Tel. (03) 3258.9565 • Fax (03) 3258.9567 Korea Tel. (02) 2278.5147 • Fax (02) 2264.1729 Malaysia Tel. (603) 5638.9940 • Fax (603) 5638.9829 Singapore Tel. 6296.7118 • Fax 6291.