Instruction Manual HI 9241 HI 93501 HI 93501S HI 93503 HI 935007 Portable Microprocessor Thermometers w w w. h a n n a i n s t .
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for correct use of the instrument, as well as a more precise idea of its versatility. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your dealer or the nearest Hanna Customer Service Center. Each meter is supplied complete with: • Temperature probe • Alkaline batteries: 3 x 1.5V AAA • Instruction manual. Note: Save all packing material until you are sure that the instrument functions correctly.
SPECIFICATIONS OF HI 9241 Main features: • • • • • HI 765 probes series Calibration Check feature Remaining battery life indication / low battery detection Auto-off capability Waterproof casing Range -50.0 to 150.0 °C Resolution 0.1 °C Accuracy ±0.4 °C excluding probe error Typical EMC Deviation ±0.5 °C (with HI 765BL probe) Battery 3 x 1.5V AAA, approx.
SPECIFICATIONS OF HI 93501 & HI 93501S Main features: • • • • • • • • HI 762 probes series Range -50 to 150 °C Calibration Check feature Remaining battery life indication / low battery detection Stability indicator (S model) Hold feature (S model) Auto-off capability Waterproof casing Range -50.0 to 150.0 °C Resolution 0.1 °C Accuracy ±0.4 °C excluding probe error Typical EMC Deviation ±0.4 °C (with HI 762PWL probe) Battery 3 x 1.5V AAA, approx.
SPECIFICATIONS OF HI 93503 Main features: • • • • • • • • HI 765 probes series Range -50 to 150 °C Calibration Check feature Remaining battery life indication / low battery detection Stability indicator Hold feature Auto-off capability Waterproof casing Range -50.0 to 150.0 °C Resolution 0.1 °C Accuracy ±0.4 °C excluding probe error Typical EMC Deviation ±0.4 °C (with HI 765PWL probe) Battery 3 x 1.5V AAA, approx.
SPECIFICATIONS OF HI 935007 Main features: • • • • • • Direct HI 766C K-thermocouple probe Available in °C Range -50.0 to 1350 °C Remaining battery life indication / low battery detection Auto-off capability Waterproof casing Range -50.0 to 199.9°C 200 to 1350 °C Resolution 0.1 °C (up to 199.9 °C) 1 °C (outside) Accuracy ±0.2% F.S. excluding probe error Typical EMC Deviation ±3 °C Battery 3 x 1.5V AAA, approx.
OPERATIONAL GUIDE Each meter is supplied with batteries. Remove the battery cap, unwrap the supplied batteries and install them while paying attention to their polarity (see also the "Battery replacement" section for more details). TAKING MEASUREMENTS Connect a suitable temperature probe to the meter (see pages 12-14 for a complete list of Hanna probes). Note: HI 935007 has a fixed probe. To switch the meter ON, press the ON/OFF button on the front of the meter.
CALIBRATION CHECK All the models are provided with automatic Cal-Check feature at start-up, which allows the user to check the instrument calibration status. If the meter is calibrated, the display will stabilize and show “0.0 °C” (±0.3 °C) in a few seconds. Contact your dealer or your nearest Hanna Service Center for recalibration, if necessary. HOLD FEATURE (HI 93501S, HI 93503) The hold function is activated by pressing the HOLD key.
WARNING MESSAGES NO PROBE message If the probe is not connected “NO PROBE” message is displayed. DAMAGED PROBE message When the readings are very far from the reading range, the probe can be damaged, “DAMAGED PROBE” message is displayed.
BATTERY REPLACEMENT When the remaining battery life is less than 10% the battery tag blinks on the display to warn the user. If the battery level is low enough to cause erroneous readings, the Battery Error Prevention System (BEPS) turns the meter off. Immediately replace the batteries with new ones. Battery replacement must only take place in a non-hazardous area. The batteries are accessed by opening the battery compartment cap on the bottom of the instrument. Replace the three 1.
TEMPERATURE PROBES Hanna offers a wide range of probes to meet your needs with air, liquid, penetration and surface measurements. All Hanna temperature probes are pre-calibrated at the factory and ready-to-use. These interchangeable probes make it possible to switch from one probe to another without requiring recalibration. Hanna probes are also available in different handle colors to avoid cross contamination during testing. Note: To clean the meters, do not use aggressive detergents.
HI 762PWL HI 762W HI 762W/10 Penetration probe w/1 m (3.3') cable & white handle Wire probe without handle for hard-to-reach places w/1 m (3.3') cable Wire probe, without handle for hard-to-reach places w/10 m (33') cable FOR HI 9241 AND HI 93503 HI 765A Air probe, w/1 m (3.3') cable & white handle HI 765A/10 Air probe, w/10 m (33') cable & white handle HI 765BL General purpose liquid probe, w/1 m (3.3') cable & black handle HI 765L General purpose liquid probe, w/1 m (3.
places w/10 m (33') cable FOR HI 935007 HI 766C (*) Penetration, K-type thermocouple probe w/1 m (3.3') cable (*) To be replaced only by authorized service technicians.
RECOMMENDATIONS FOR USERS Before using these products, make sure they are entirely suitable for the environment in which they are used. Operation of these instruments in residential areas could cause unacceptable interferences to radio and TV equipment, requiring the operator to follow all necessary steps to correct interferences. Any variation introduced by the user to the supplied equipment may degrade the instruments’ EMC performance.
Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA Technical Support for Customers Tel. (800) 426 6287 Fax (401) 765 7575 E-mail tech@hannainst.com www.hannainst.