Instruction Manual HI 9041 - HI 9061 HI 93501 - HI 93502 HI 93503 HI 935007 Portable Microprocessor Thermometers w w w . h a n n a i n s t .
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for correct use of the instrument, as well as a more precise idea of its versatility. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. These instruments are in compliance with the directives.
PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your dealer or the nearest Hanna Customer Service Center. Each meter is supplied complete with: • Temperature probe (for HI 93501 and HI 93503 only); • Alkaline batteries: 1 x 9V for HI 9041; 4 x 1.5V AA for HI 9061; 3 x 1.5V AA for HI 93501, HI 93502, HI 93503 and HI 935007; • Instruction manual.
SPECIFICATIONS OF HI 9041C & HI 9041F Main features: • • • • HI 765 probes series Available HI 765 accuracy test plugs Remaining battery life indication / low battery detection Available in °C or °F models Range Resolution Accuracy Typical EMC Deviation Probe Battery Environment Dimensions Weight HI 9041 -50.0 to 150.0°C (HI 9041C) -58.0 to 302.0°F (HI 9041F) 0.1°C / 0.1°F ±0.4°C / ±0.8°F excluding probe error ±0.5°C / ±1°F with HI 765BL probe HI 765 series 1x9V (6LR61) / approx.
SPECIFICATIONS OF HI 9061C & HI 9061F Main features: • • • • • HI 765 probes series Available HI 765 accuracy test plugs Remaining battery life indication / low battery detection Available in °C or °F models Waterproof casing Range Resolution Accuracy Typical EMC Deviation Probe Battery Environment Dimensions Weight HI 9061 -50.0 to 150.0°C (HI 9061C) -58.0 to 302.0°F (HI 9061F) 0.1°C / 0.1°F ±0.4°C / ±0.8°F excluding probe error ±0.5°C / ±1°F with HI 765BL probe HI 765 series 4x1.5V AA (LR6) / approx.
SPECIFICATIONS OF HI 93501 Main features: • • • • • • • HI 762 probes series Available HI 762 accuracy test plugs Range -50 to 150°C Calibration Check feature Remaining battery life indication / low battery detection Auto-off capability Waterproof casing SPECIFICATIONS Range -50.0 to 150.0°C Resolution 0.1°C Accuracy ±0.4°C excluding probe error Typical EMC Deviation ±0.4°C (with HI 762PWL probe) Battery 3x1.
SPECIFICATIONS OF HI 93502 Main features: • • • • • • • LUMBERG® socket for temperature probes LUMBERG® probes series Range -50 to 150°C Calibration Check feature Remaining battery life indication / low battery detection Auto-off capability Waterproof casing SPECIFICATIONS Range -50.0 to 150.0°C Resolution 0.1°C Accuracy ±0.4°C excluding probe error Typical EMC Dev. ±0.4°C (with HI 762PW/LUM probe) Battery 3x1.
SPECIFICATIONS OF HI 93503 Main features: • • • • • • • HI 765 probes series Available HI 765 accuracy test plugs Range -50 to 150°C Calibration Check feature Remaining battery life indication / low battery detection Auto-off capability Waterproof casing SPECIFICATIONS Range -50.0 to 150.0°C Resolution 0.1°C Accuracy ±0.4°C excluding probe error Typical EMC Deviation ±0.4°C (with HI 765PWL probe) Battery 3x1.
SPECIFICATIONS OF HI 935007 Main features: • • • • • • Direct HI 766C K-thermocouple probe Measuring scale selection (°C or °F) Range -50.0 to 900°C or -58.0 to 1650°F Remaining battery life indication / low battery detection Auto-off capability Waterproof casing SPECIFICATIONS Range -50.0 to 199.9°C / 200 to 900°C -58.0 to 399.9°F / 400 to 1650°F Resolution 0.1°C (up to 199.9°C) / 1°C (outside) 0.1°F (up to 399.9°F) / 1°F (outside) Accuracy (@20°C/68°F) ±0.2% F.S.
OPERATIONAL GUIDE Each meter is supplied complete with batteries. Remove the back cover, unwrap the supplied batteries and install them while paying attention to their polarity (see also the "Battery replacement" section for more details). Note: HI 93501, HI 93502, HI 93503 and HI 935007 come with batteries already installed. TAKING MEASUREMENTS Connect a suitable temperature probe to the meter (see pages 14-15 for a complete list of Hanna probes). Note: HI 935007 has a fixed probe.
Reattach the back making sure that the gasket is in place and tighten the 4 screws to ensure a watertight seal. CALIBRATION CHECK (for HI 93501, HI 93502 and HI 93503) The HI 93501, HI 93502 and HI 93503 models are provided with the Cal-Check feature, which allows the user to check the instrument calibration status at any time. Press and hold the CAL CHECK button. If the meter is calibrated, the display will stabilize and show “0.0°C” (±0.3°C) in a few seconds.
using the battery type specified in this instruction manual. • HI 9041 and HI 9061: remove the battery cover and replace the batteries (1x9V for HI 9041 or 4x1.5V AA for HI 9061) with new ones, while paying attention to their polarity. Reattach the battery cover. • HI 93501, HI 93502, HI 93503 and HI 935007: the batteries are accessed by separating the front and the back halves of the meter. Unscrew the 4 screws on the back of the meter and carefully replace the three 1.
FACTORY RECALIBRATION All Hanna thermometers have been accurately pre-calibrated at the factory. It is generally recommended to have all thermometers recalibrated at least once a year. For an accurate recalibration, contact your nearest Hanna Customer Service Center. Hanna Accuracy Test Plugs provide a quick and easy way to check the meter's accuracy by simply connecting these Test Plugs to the probe connector of the meter. If the reading differs by more than ±0.4°C (±0.
TEMPERATURE PROBES Hanna offers a wide range of probes to meet your needs with air, liquid, penetration and surface measurements. All Hanna temperature probes are pre-calibrated at the factory and ready-to-use. These interchangeable probes make it possible to switch from one probe to another without requiring recalibration. Hanna probes are also available in different handle colors to avoid cross contamination during testing.
FOR HI 93502 (probes with Lumberg® connector) HI 762A/LUM Air probe, w/1 m (3.3') cable & white handle HI 762BP/T/LUM Frozen Food between pack probe, w/1 m (3.3') cable & white handle HI 762L/LUM General purpose liquid probe, w/1 m (3.3') cable & white handle HI 762PW/LUM General purpose penetration probe, w/1 m (3.
MANTHFOODR1 06/03 w w w . h a n n a i n s t .