Instruction Manual HI 935005 - HI 935005N HI 935002 - HI 935009 Portable K-Thermocouple Thermometers w w w . h a n n a i n s t .
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for correct use of the instrument, as well as a more precise idea of its versatility. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. These instruments are in compliance with the directives.
SPECIFICATIONS OF HI 935005 & HI 935005N Range (*) Resolution Accuracy (@20°C/68°F) Typical EMC Deviation Battery Auto-off (HI 935005 only) Environment Dimensions Weight SPECIFICATIONS -50.0 to 199.9°C / 200 to 1350°C -58.0 to 399.9°F / 400 to 2462°F 0.1°C (up to 199.9°C) / 1°C (outside) 0.1°F (up to 399.9°F) / 1°F (outside) ±0.2% full scale for one year, excluding probe error ± 3°C / ± 6°F with HI 766 K-thermocouple probe 3x1.5V AA (IEC LR6) batteries, approx.
SPECIFICATIONS OF HI 935002 & HI 935009 HI 935002 Range (*) Resolution Accuracy (@20°C/68°F) Typical EMC Deviation Battery Auto-off (HI 935009 only) Environment Dimensions Weight HI 935009 SPECIFICATIONS -50.0 to 199.9°C / 200 to 1350°C -58.0 to 399.9°F / 400 to 2462°F 0.1°C (up to 199.9°C) / 1°C (outside) 0.1°F (up to 399.9°F) / 1°F (outside) ±0.2% full scale for one year, excluding probe error ± 3°C / ± 6°F with HI 766 K-thermocouple probe 3x1.5V AA (IEC LR6) batteries, approx.
OPERATIONAL GUIDE To switch the instrument ON, press the ON/OFF key. The thermometer will carry out a self diagnostic test routine, the LCD will show all segments for a few seconds (or as long as ON/OFF is held), followed by the percentage indication of the remaining battery life. The thermometer then enters normal measurement mode. If a temperature probe is plugged in, the meter displays the measured temperature.
HI 935009 allows the user to select the auto-off time period through the front keyboard; press (ALT+) AOFF to enter the mode, then set the desired auto shut-off interval (8 min, 60 min, or disabled) with the ALT key; release all buttons, wait a few seconds and the meter will return to normal measurement mode.
USER CALIBRATION BATTERY REPLACEMENT The HI 935005N and HI 935009 models can be calibrated at 0°C by using an ice bath. • Prepare an ice bath with approximately equal volumes of distilled water and chopped ice made from distilled water. • Immerse the temperature probe in the center of the ice bath, taking care not to touch the ice with the probe tip. • Ensure that the meter is measuring a temperature within ±3°C. • To enter the Calibration mode, press and hold the CAL button for about 5 seconds.
ACCESSORIES K-TYPE THERMOCOUPLE PROBES OTHER ACCESSORIES with integral handle, 1 m (3.
SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 Indonesia: Tel. (21) 4584.2941 • Fax (21) 4584.2942 Japan: Tel. (03) 3258.9565 • Fax (03) 3258.9567 Korea: Tel. (02) 2278.5147 • Fax (02) 2264.1729 Malaysia: Tel. (603) 5638.9940 • Fax (603) 5638.9829 Singapore: Tel. 6296.