User Guide

9
HOLD mode.
Note: Although the display is frozen, internally the meter continues
measuring and updating Hi/Lo and relative values.
In HI 935009 only, the held value is also stored in non-volatile
memory and can be recalled by pressing (ALT+) MR key. While the
MR key is held, the "HOLD" tag lights up and the meter displays the
stored value; when the MR key is released, after 1 second the meter
returns to normal measurement mode.
HIGH/LOW TEMPERATURES
The maximum and minimum tempera-
tures are continuously monitored and
displayed on the lower portion of the LCD.
Note: When the reading goes over-range
or the probe is removed, the Hi and
Lo values display dashes until cleared.
CLEAR FUNCTION
Upon pressing the CLR key, the High/Low values may be cleared at
any time during measurement and the current reading is assigned to
the highest and lowest temperature values for the displayed channel
only.
AUTO SHUT-OFF (HI 935005 and HI 935009)
To save battery life, the HI 935005 and HI 935009 models are
provided with an auto-off feature, which switches the instrument off
after a certain period of nonuse.
The delay time can be changed by the user (8 minutes, 60 minutes,
or disabled).
To access this feature, the HI 935005 is provided with an internal
button located in the battery compartment (see figure on page 10).
Unscrew the 4 screws on the back of the meter and select the desired
auto-off time period by pressing the button and viewing the value on
the LCD. Re-attach the back making sure that the gasket is in place
and tighten the 4 screws to ensure a watertight seal.