User Guide

6
SPECIFICATIONS OF
HI 935002 & HI 935009
Main features:
Two independent measurement channels, T1 and T2
HI 766 K-thermocouple probes
Display of temperature in degrees Celsius or Fahrenheit
Range -50 to 1350°C or -58 to 2462°F
HOLD function
HI and LO (Max and Min) temperature values always displayed
on the LCD
DISP function to select the information to be displayed (Normal,
Relative or T1/T2)
Remaining battery life indication / low battery detection
Waterproof casing
Auto-off capability (HI 935009 only)
MR function for recalling memorized reading (HI 935009 only)
Backlight feature (HI 935009 only)
User calibration at 0°C (HI 935009 only)
SPECIFICATIONS
Range (*) -50.0 to 199.9°C / 200 to 1350°C
-58.0 to 399.9°F / 400 to 2462°F
Resolution 0.1°C (up to 199.9°C) / 1°C (outside)
0.1°F (up to 399.9°F) / 1°F (outside)
Accuracy ±0.2% full scale
(@20°C/68°F) for one year, excluding probe error
Typical EMC ± 3°C / ± 6°F
Deviation with HI 766 K-thermocouple probe
Battery 3x1.5V AA (IEC LR6) batteries,
approx. 1600 hours of continuous use (BL off)
Auto-off user selectable: 8 min, 60 min, disabled
(HI 935009 only)
Environment -10 to 50°C (14 to 122°F); RH 100%
Dimensions 150 x 80 x 36 mm (5.9 x 3.1 x 1.4")
Weight 235 g (8.3 oz.)
(*) Range may be limited by probe.