User Guide

12
USER CALIBRATION
The HI 935005N and HI 935009 models can be calibrated at 0°C by
using an ice bath.
Prepare an ice bath with approximately equal volumes of distilled
water and chopped ice made from distilled water.
Immerse the temperature probe in the center of the ice bath,
taking care not to touch the ice with the probe tip.
Ensure that the meter is measuring a temperature within ±3°C.
To enter the Calibration mode, press and hold the CAL button for
about 5 seconds.
The CAL tag turns on to indicate that
the Calibration mode has been entered.
Note: If the measurement is outside the
±3°C window, the meter does not
enter Calibration mode.
When the meter reaches the stability condition, which is detected
when the measurement remains within ±0.2°C for 5 seconds,
the calibration is accepted and the reading becomes 0°C (32°F).
The meter then automatically returns to normal mode.
Note: To exit the Calibration mode at any time, press the CAL key.
Note: User calibration cannot be entered in relative or in T1-T2 mode.
Note: User calibration is only performed on the current channel
displayed (T1 or T2).
All Hanna thermometers have been accurately pre-calibrated at the
factory.
It is generally recommended to have all thermometers recalibrated at
least once a year.
For an accurate recalibration, contact your nearest Hanna Customer
Service Center.