User Guide
12
USER CALIBRATION
The HI 935005N and HI 935009 models can be calibrated at 0°C by
using an ice bath.
• Prepare an ice bath with approximately equal volumes of distilled
water and chopped ice made from distilled water.
• Immerse the temperature probe in the center of the ice bath,
taking care not to touch the ice with the probe tip.
• Ensure that the meter is measuring a temperature within ±3°C.
• To enter the Calibration mode, press and hold the CAL button for
about 5 seconds.
• The CAL tag turns on to indicate that
the Calibration mode has been entered.
Note: If the measurement is outside the
±3°C window, the meter does not
enter Calibration mode.
• When the meter reaches the stability condition, which is detected
when the measurement remains within ±0.2°C for 5 seconds,
the calibration is accepted and the reading becomes 0°C (32°F).
• The meter then automatically returns to normal mode.
Note: To exit the Calibration mode at any time, press the CAL key.
Note: User calibration cannot be entered in relative or in T1-T2 mode.
Note: User calibration is only performed on the current channel
displayed (T1 or T2).
All Hanna thermometers have been accurately pre-calibrated at the
factory.
It is generally recommended to have all thermometers recalibrated at
least once a year.
For an accurate recalibration, contact your nearest Hanna Customer
Service Center.
FACTORY RECALIBRATION