Manual

29
If everything is satisfactory
the LCD will blink "50"
expecting the second
buffer (50°C/122°F).
Note: if a single point cali-
bration is required,
press the CAL key
to leave the calibra-
tion mode now.
Place the probe into the
second container with hot
water (50°C/122°F).
When "50" stops blinking, press the CFM
key again to confirm the second calibra-
tion point.
If the value is out of the slope range, an
error message will show on the secondary
display "E2". Press the CAL key to re-
calibrate the value.
If everything is satisfactory the tempera-
ture calibration is complete and the meter
will return to the normal operational mode.
CAL
CFM
CAL
°C
50 °C
(122 °F)