Manual

28
into the PVC sleeve with the holes towards
the top (the end nearest to the cable) of
the probe.
Place the probe into the
container with ice and wa-
ter (0.0°C/32°F).
Press the CAL key. The secondary (lower)
display will blink "0" for about 30 sec-
onds.
When "0" stops blinking, the calibration
can be confirmed. Press the CFM key to
confirm the first value.
If the value is out of the offset range, an
error message will show on the secondary
display "E1". Press the CAL key to re-
calibrate the value.
CAL
CFM
CAL
0 °C
(32 °F)
°
C