Manual

20
PROCEDURE FOR SINGLE POINT
CALIBRATION WITH BUFFER SOLUTION
HI933300 and HI933301 have auto-buffer rec-
ognition: you can simply place the probe into
the type of buffer solution that you need to
calibrate and perform the single point calibra-
tion very quickly as described below.
Choose one of the following calibration solu-
tions:
84.0 µS/cm (cal. solution HI 7033/HI8033)
1413 µS/cm (cal. solution HI 7031/HI8031)
12.88 mS/cm(cal. solution HI 7030/HI8030)
80.0 mS/cm(cal. solution HI 7034/HI8034)
Fill a beaker with conduc-
tivity calibration solution
(e.g. HI 7030 or HI8030,
12.88 mS/cm).
Press the RANGE key to
select the conductivity
measurement mode.
Dip the probe into the
sample. The level of solu-
tion must be higher than
the holes on the PVC
sleeve. Tap the probe re-
peatedly on the bottom of
the beaker and stir it to ensure that no air
bubbles are trapped inside the sleeve.
Press the CAL key and the buffer value
will be blinking on the display (e.g.
12.88 mS/cm).
8 cm
(3¼")
RANGE
CAL
mS