PROCEDURE FOR OFFSET CALIBRATION
(at 0.0 µS/cm)
To perform the Offset
calibration, dry the con-
ductivity probe and leave
it in air.
Press the ALT and the CAL keys simulta-
neously. The primary LCD display will blink
"0.0 µS" for about 10 seconds.
When "0.0 µS" stops blinking, the cali-
bration can be confirmed. Press the CFM
key to confirm the offset value.
If the input reading is out of the offset
range, an error message will show on the
secondary display "E0". Press the CAL
key to re-calibrate the value.
If everything is satisfactory
the LCD will blink
"84.0 µS", expecting the
first calibration solution.
Press the CAL key to exit
the calibration mode now.
CAL
ALT
µS
µS
CFM
CAL
CAL
µS