User Guide

15
15
The instrument should be re-calibrated at least once a month, or
when the probe or battery is changed.
Note: For more accurate results, it is advisable to use a calibration
solution close to the range to be measured. See the accessories
section on page 23 for a wide selection of conductivity solu-
tions.
PROCEDURE FOR HI 8734
Pour sufficient quantity of a TDS calibra-
tion solution (e.g. HI 7032) into a beaker
to cover the holes on the probe. If pos-
sible, use plastic beakers to minimize any
EMC interference.
Immerse the TDS probe, making sure
that holes are completely submerged, and
the ChecktempC in the solution.
Wait for a couple of minutes for thermal
equilibrium to be reached.
Tap the probe on the bottom, then shake it while rotating to make
sure no air bubbles remain trapped in the sleeve.
Record the temperature of the HI 7032 buffer solution from the
ChecktempC (e.g. 18°C).
Switch the instrument on by pressing
ON/OFF.
Press TDS/TEMP to display the temperature.
Adjust the TEMPERATURE knob to display 18°C.
Press TDS/TEMP again to display the TDS
measurement.
Select 1999 mg/L range by pressing the ap-
propriate range key.
H
I 7
0
3
0
°C
ON
OFF
TDS
TEMP.
TDS
TEMP.
TEMPERATURE
1999
mg/l
Untitled-19
16
16
Adjust the calibration trimmer (see page 7,#6) on the side of the
instrument with the calibration screwdriver until the display shows
the TDS reading at 25°C (see the TDS vs. temperature chart on
page 19). E.g. @25°C "1382 mg/L".
All subsequent measurements will be compensated to 25°C (77°F).
If you prefer to standardize the temperature compensation to
20°C (68°F) rather than 25°C (77°F), leave the TEMPERATURE
knob to 18°C (if the temperature of the solution is 18°C). Adjust
the trimmer to read "1251 mg/L" (see the TDS vs. temperature
chart on page 19). All subsequent measurements will be compen-
sated to 20°C.
The calibration is now complete and the instrument is ready for
use.
The instrument should be re-calibrated once a month, or when the
probe or battery is changed.
Note: For more accurate results, it is advisable to use a calibration
solution close to the range to be measured. See the accessories
section on page 23 for a wide selection of TDS solutions.
PROCEDURE FOR HI 933000
Pour sufficient quantity of a conductivity
calibration solution (e.g. HI 7030/
HI 8030) into a beaker to cover the holes
on the probe. If possible, use plastic bea-
kers to minimize any EMC interference.
Immerse the conductivity probe in the
solution, making sure that holes are com-
pletely submerged.
Wait for a couple of minutes for thermal
equilibrium to be reached.
Tap the probe on the bottom, then shake it while rotating to make
sure no air bubbles remain trapped in the sleeve.
T
E
M
P
E
R
A
T
U
R
E
HI 7030