User Guide

19
19
The TDS value in aqueous solutions is directly proportional to conduc-
tivity. The ratio between the two parameters depends on the solution
and usually it is set to a factor of 0.5 (corresponding to a solution of
CaCO
3
). This means that 1 µS/cm is equal to 0.5 mg/L (ppm) of TDS.
For manual temperature compensation, refer to the following chart:
TDS VERSUSTDS VERSUS
TDS VERSUSTDS VERSUS
TDS VERSUS
TEMPERATURE CHARTTEMPERATURE CHART
TEMPERATURE CHARTTEMPERATURE CHART
TEMPERATURE CHART
For instance, the TDS values of the calibration solutions at 25°C are
1382 mg/L or 12.41 g/L when using HI 7032 or HI 7036, respec-
tively.
At 20°C, the values are 1251 mg/L or 11.24 g/L, respectively.
With the solutions at 30°C, the values are 1515 mg/L or 13.61 g/L,
respectively.
°C °F HI 7032 HI 7036
mg/L g/L
(ppm) (ppt)
0 32 758 6.82
5 41 876 7.88
10 50 999 8.99
15 59 1122 10.10
16 60.8 1148 10.33
17 62.6 1173 10.56
18 64.4 1200 10.78
19 66.2 1224 11.01
20 68 1251 11.24
21 69.8 1277 11.47
22 71.6 1303 11.71
23 73.4 1329 11.94
24 75.2 1358 12.18
2525
2525
25
7777
7777
77
13821382
13821382
1382
12.4112.41
12.4112.41
12.41
26 78.8 1408 12.65
27 80.6 1438 12.89
28 82.4 1461 13.13
29 84.2 1476 13.37
30 86 1515 13.61
31 87.8 1541 13.85
Untitled-19
20
20
Highly acidic, alkaline samples or solutions with high salt content
might have a different coefficient than the customary 2% per degree °C.
In order to calculate this coefficient follow the procedure below:
Immerse the probe of HI 8733 in the sample and adjust the
TEMPERATURE COEFFICIENT knob to 0% (i.e. no compensation).
Condition the sample and probe to 25°C and note the conductiv-
ity reading, C
25
.
Condition the sample and probe to a different temperature t°C
(approximately 10°C different from 25°C) and note the conduc-
tivity reading C
t
.
The temperature coefficient ß of the solution is calculated as given
by the following formula:
(C
t
- C
25
)
ß = 100 x
(t - 25) x C
25
The above procedure is suitable for determining the temperature
coefficient in a laboratory or where the temperature of the solution
can be controlled.
If this is not possible (e.g. on-site measurements), the following
procedure can be used providing the sample temperature varies by at
least 5°C or preferably 10°C:
Immerse the probe of HI 8733 in the test solution and turn the
TEMPERATURE COEFFICIENT knob to 0% (no compensation).
Check the conductivity reading and record the value. Make sure
the reading is stable, i.e. no greater variations than ±0.2 mS/
cm within a minute.
Repeat the procedure when the temperature of the test solution
has changed by at least 5°C. Wait for the conductivity reading to
stabilize.
Adjust the TEMPERATURE COEFFICIENT knob until the display
shows the same value as recorded earlier.
The value indicated by the knob is the temperature coefficient of
the solution.
2.5
2
1.51
0.5
0
TEMPERATURE COEFFICIENT
DETERMINING THE TEMPERATUREDETERMINING THE TEMPERATURE
DETERMINING THE TEMPERATUREDETERMINING THE TEMPERATURE
DETERMINING THE TEMPERATURE
COEFFICIENT OF A SOLUTION (HI 8733)COEFFICIENT OF A SOLUTION (HI 8733)
COEFFICIENT OF A SOLUTION (HI 8733)COEFFICIENT OF A SOLUTION (HI 8733)
COEFFICIENT OF A SOLUTION (HI 8733)