SPECIFICATIONS HI 931410
Range pH
ISE
ORP
°C
Resolution pH
ISE
ORP
°C
Accuracy pH
(@ 20°C / 68°F) ISE
ORP
°C
Typical EMC pH
DeviationISE/ORP
°C
Recorder pH
Output m V
Max. Current
Offset
Slope
Calibration
Temperature
Compensation
Electrode
Temp. Probe
Input Impedance
Power
Environment
Dimensions
Shipping Weight
0.00 to 14.00
0.0 to ±399.9mV
±1999mV
-10.0° to 125.0
0.01
0.1mV
0.1/1mV
0.1
±0.01
±0.2 mV
±1 mV
±0.5
±0.01
±0.2mV
±0.7
0 to 2 V
0 to 2 V
20mA
±1 pH
from 85% to 105%
Automatic 1 or 2 points with
3 memorized buffers
(pH 4.01, 7.01, 10.01)
Manual from 0 to 100°C (32 to 122°F)
Automatic from -10 to 125°C (14 to 257°F)
HI 1131B glass body combination
refillable pH electrode (included)
HI 7669/2W (included)
10
12
Ω
12VDC (CSA, UL & CE approved)
0 to 50°C (32 to 122°F); 95% RH
230x170x70 mm (9.1x6.7x2.7")
1.5 Kg (3.3 lb.)