Immerse the temperature probe
in the vessel with hot water.
When the temperature reading
has stabilized, use the
°C
and
°C
keys to set the reading on
the secondary LCD to the
noted temperature of the hot
water.
Press CFM key.
The primary LCD
section will show
"390.0 mV".
Complete your temperature calibra-
tion by pressing the CAL key if
you do not want to perform the mV
calibration mode.
mV CALIBRATION OF HI 931402
Initial Preparation
Prepare HI 931001 or a simulator that can
provide ±390.0 mV with accuracy of ±0.1mV
and ±1000 mV with accuracy of ±1 mV.
Procedure
Continue from the tempera-
ture calibration, the primary
LCD section will show
"390.0 mV".
Select 390.0 mV of the simulator.
Wait for 3 sec., press the CFM
key. The primary LCD will display
the next reading to be selected on
the simulator.
Repeat previous two steps for -390.0 mV,
1000 mV and -1000 mV.
At the end of the mV calibration, the meter
will return to the pH measurement mode.