User Guide

19
TWO POINT CALIBRATION
A two point calibration is recommended for
best accuracy. As a second point use pH 4.01
if you are going to measure acid samples (pH 7
or less), use pH 9.18 or pH 10.01 if you are going
to measure alkaline samples.
Two point calibration at pH 6.86 and pH 4.01
or pH 9.18 (NBS):
Proceed as described in “One point calibra-
tion at pH 6.86” but do not quit calibration by
pressing the CAL key at the end.
After the first calibration point
is confirmed, immerse the pH
electrode and the temperature
probe into the second buffer
solution (pH 4.01 or pH 9.18)
and shake briefly.
Note: the electrode should
be submerged approxi-
mately 4 cm (1½") into the
solution. The temperature
probe should be located as
close as possible to the pH electrode.
Select the second buffer
value on the secondary
display by pressing the
°C
or
°C
key.
When the “READY” and
“CON” symbols blink on the
display the reading is stable
and the calibration can be
confirmed.
Press the CFM key. If the
reading is not close to the
selected buffer solution,
WRONG
BU
F
2
” and “WRONG
will blink alternatively. If
the reading is close to the
selected buffer, the slope and
the offset are calibrated.
°C
HI 7004
4 cm
(1½
")
ºC
READY
CON
CFM
WRONG
WRONG
2
BUF