User Guide

17
Only when the reading is
stable, "READY" and "CON"
will blink.
Press the CFM key to con-
firm the calibration. If the
reading is not close to the
selected buffer, "WRONG
BUF
1
"
and "WRONG " will blink
alternatively. If the reading
is close to the selected
buffer, the meter stores the
reading (and adjusts the off-
set point). The buffer value is
then displayed on the primary
LCD and the secondary LCD
will display "4.01".
Press the CAL key and the
calibration process is ended
with only the offset of the
meter calibrated.
For best accuracy however, it is recommended
that a two point calibration is performed (see
page 19).
One point calibration at pH 7.01 (STAN-
DARD):
Immerse the pH electrode
and the temperature probe
into pH 7.01 buffer solution
and shake briefly.
Note: the electrode should be
submerged approximately
4 cm (1½") into the solution.
The temperature probe should
be located as close as pos-
sible to the pH electrode.
Press the CAL key. The “CAL” and “
BUF
1
indicators will be displayed. The second-
ary LCD will display “4.01”.
CAL
1
BUF
pH
READY
CON
CFM
CAL
CAL
1
BUF
pH
WRONG
WRONG
1
BUF
HI 7007
4 cm
(1½
")
CAL