User Guide

21
When "0" stops blinking, the calibration
can be confirmed. Press the CFM key to
confirm the first temperature value.
If the value is out of the offset range, "E1"
as error message will show on the second-
ary display. Press the CAL key to re-cali-
brate the value.
If everything is satisfactory
the secondary LCD will
blink "50" expecting the
second temperature mea-
surement.
Note: If a single point cali-
bration is required,
press the CAL key
to leave the tempera-
ture calibration mode
now. However, it is
always better to pro-
ceed as follows for a
two-point calibration.
• Immerse the temperature probe
in the container at 50°C
(122°F). Place the probe as
close as possible to the refer-
ence thermometer.
When "50" stops blinking, indicating that
the probe has stabilized, press the CFM
CFM
CAL
CAL
50 °C
(122 °F)
°C