User Guide
21
• When "0" stops blinking, the calibration
can be confirmed. Press the CFM key to
confirm the first temperature value.
• If the value is out of the offset range, "E1"
as error message will show on the second-
ary display. Press the CAL key to re-cali-
brate the value.
• If everything is satisfactory
the secondary LCD will
blink "50" expecting the
second temperature mea-
surement.
Note: If a single point cali-
bration is required,
press the CAL key
to leave the tempera-
ture calibration mode
now. However, it is
always better to pro-
ceed as follows for a
two-point calibration.
• Immerse the temperature probe
in the container at 50°C
(122°F). Place the probe as
close as possible to the refer-
ence thermometer.
• When "50" stops blinking, indicating that
the probe has stabilized, press the CFM
CFM
CAL
CAL
50 °C
(122 °F)
°C