TEMPERATURE CALIBRATION
(for Technical Personnel only)
You can perform the temperature calibration
as follows.
INITIAL PREPARATION
Prepare a vessel containing
ice and water and another
one containing hot water (at
a temperature of 50°C/122°F). Place insu-
lation material around the container to mini-
mize temperature changes.
Use a ChecktempC or a calibrated ther-
mometer with a resolution of 0.1°C as a
reference thermometer.
PROCEDURE
Turn the meter ON and
press the RANGE key twice
to get just the temperature
reading on the display.
Immerse the temperature probe
in the container of ice and wa-
ter. Place the probe as close
as possible to the reference ther-
mometer.
Press the CAL key. The secondary display
(lower) LCD will blink "0" for approximately
30 seconds.
0 °C
(32 °F)
°
C
50 ºC
(122 ºF)
°C
ON/OFF
RANGE
°C
0 °C
(32 °F)
°C
CAL