User Guide

17
confirm the second calibration point.
If the buffer is out of the slope range, "E2"
as error message will appear on the sec-
ondary display. Press the CAL key and
recalibrate the value.
If everything is satisfactory
the secondary LCD will blink
"4" (if previously was "10")
or "10" (if previously was
"4") expecting the third
buffer (HI7004/HI8004 or
HI7010/HI8010).
Note: if a two-point calibra-
tion is required, press
the CAL key to leave
the calibration mode
now.
Rinse the electrode with some pH 4.01 (or
pH 10.01) solution. Dip the bottom 4 cm
(1½") of the electrode in a beaker contain-
ing the third buffer.
When "4" (or "10") stops blinking, indicat-
ing that the electrode has stabilized, press
the CFM key again to confirm the third
CFM
CAL
CAL
4 cm
(1½
")
HI 7004