User Guide

15
PROCEDURE
Press the RANGE key to select the pH
measurement mode.
Rinse the pH electrode with
some pH 7.01 (HI7007/
HI8007) solution or tap wa-
ter. Dip the bottom 4 cm
(1½") of the electrode and
the temperature probe into
a beaker containing pH 7.01
solution.
Press the CAL key. The secondary (lower)
LCD will blink “7” to indicate that pH 7.01
is expected and the electrode is getting
conditioned.
When the “7” stops blinking, the calibration
can be confirmed. Press the CFM key to
confirm the first buffer.
• If everything is satisfactory
the secondary LCD will blink
10” expecting the second
buffer at pH 10.01.
If the buffer is out of the offset range, "E1"
as error message will appear on the sec-
ondary display. Press the CAL key and
RANGE
pH
4 cm
(1½
")
HI 7007
CAL
CFM