User Guide

17
17
If the reading is not close
to the selected buffer, the
"BUF" prompt will continue
to blink.
If on the other hand, the
reading is close to the se-
lected buffer, the meter
will recognize the standard
solution. The BUF symbol
will stay still on the dis-
play, while the "
" sym-
bol will blink, indicating that
the meter is stabilizing its
reading.
• Once stabilized "CON"
symbol will appear on
the display.
Wait for a couple of min-
utes and press the CFM
key.
The meter will return to its normal mode
and the two point calibration process is
completed.
CFM
mannacl_r1.P65
18
18
NaCl CALIBRATION (HI931100/HI931102)
The calibration is performed by immersing
the FC 300B sodium-sensitive electrode in 2
different standard NaCl solutions to achieve
accurate 2 point calibration.
It is also possible to calibrate just the offset
(3.00 g/L NaCl) following the single point cali-
bration procedure described below.
For best accuracy, a two-point calibration is
recommended (see below).
PREPARATION
For HI931100 and HI 931102:
Pour small quantities of
HI 7083 or HI 8083 (3.00 g/L
NaCl) solution and HI 7085
or HI 8085 (0.30 g/L) or
HI 7081 or HI 8081 (30 g/L)
into two clean beakers.
To get accurate readings use HI7083/HI 8083
and HI 7085/HI8085 if you are going to mea-
sure samples with low sodium chloride con-
tent (below 3 g/L) or HI 7083/HI8083 and
HI 7081/HI8081 if you are going to measure
samples with high sodium chloride content
(over 3 g/L).
For accurate calibra-
tion, use two bea-
kers for each buffer
calibration, the first
one for rinsing the electrode, the second
one for calibration. This way, contamina-
tion of the buffer is mini-
mized. Whenever possible
use plastic beakers to mini-
mize any EMC interferences.
• Turn the meter on by
pressing the ON/OFF key.
Remove the protective cap
of the electrode and rinse
HI 7083
HI 7085
H
I
7
0
8
3
ON/OFF
ON/OFF