User Guide

9
9
Measurement
NaCl
NaCl %
°C
Resolution
NaCl
NaCl %
°C
Accuracy NaCl
(@20°C/68°F) NaCl%
°C
Typical NaCl
EMC NaCl%
Deviation °C
Calibration
Temperature
Compensation
Electrode
Temperature
Probe
Battery Type
Life
Environment
Dimensions
Weight
0.150 to 1.500 g/L
1.50 to 15.00 g/L
15.0 to 150.0 g/L
150 to 300 g/L
0.00 to 30.00
0.0 to 80.0
0.001 g/L
0.01 g/L
0.1 g/L
1 g/L
0.01
0.1
±5% of reading
±5% of reading
±0.5
±2% F.S.
±2% F.S.
±0.5
Automatic 1 or 2 point
at 3.00 g/L (HI 7083)
and 0.30 g/L (HI 7085)
or 30.0 g/L (HI 7081)
fixed at 25°C (77°F)
FC 300B glass combination
sodium-sensitive electrode
(not included)
HI 7662 temperature probe
(not included)
4 x 1.5V AA size
200 hours of continuous use
0 to 50°C (32 to 122°F);
100% RH
196 x 80 x 60 mm
(7.7 x 3.1 x 2.4")
450 g (15.8 lb.)
SPECIFICATIONS HI 931102
HI 931102
mannacl_r1.P65
10
10
OPERATIONAL GUIDE
INITIAL PREPARATIONINITIAL PREPARATION
INITIAL PREPARATIONINITIAL PREPARATION
INITIAL PREPARATION
Each meter is supplied
complete with 4x1.5V AA
batteries.
Remove the back cover,
unwrap the batteries and
install them while paying
attention to their polarity.
PROBES CONNECTIONPROBES CONNECTION
PROBES CONNECTIONPROBES CONNECTION
PROBES CONNECTION
Connect the sodium-sen-
sitive electrode
FF
FF
F
CC
CC
C
300B300B
300B300B
300B
to the BNC socket on
the top of the instrument.
Connect the optional tem-
perature probe
HH
HH
H
II
II
I
7662 7662
7662 7662
7662 to
the RCA socket.
The temperature probe can be used indepen-
dently to take temperature measurements.
If the probe is
not plugged in,
the secondary
display reads
25°C and the
"°C" indicator
blinks.
To switch the instrument on,
press and hold the ON/OFF
key for a fraction of a second.
These meters have a built-in
protection against electromag-
netic interferences and the
holding of the keys is to en-
FC 300B
SODIUM
ELECTRODE
HI 7662N
TEMPERATU
RE
PROBE
gm
°C
NaCl
ON/OFF
ON/OFF