5
5
Measurement
NaCl
°C
Resolution
NaCl
°C
Accuracy NaCl
(@20°C/68°F) °C
Typical NaCl
EMC Deviation °C
Calibration
Temperature
Compensation
Electrode
Temperature
Probe
Battery Type
Life
Environment
Dimensions
Weight
SPECIFICATIONS HI 931100
HI 931100
0.150 to 1.500 g/L
1.50 to 15.00 g/L
15.0 to 150.0 g/L
150 to 300 g/L
0.0 to 80.0°C
0.001 g/L
0.01 g/L
0.1 g/L
1 g/L
0.1°C
±5% of reading
±0.5°C
±2% F.S.
±0.5°C
Automatic 1 or 2 point
at 3.00 g/L (HI 7083)
and 0.30 g/L (HI 7085)
or 30.0 g/L (HI 7081)
fixed at 25°C (77°F)
FC 300B glass combination
sodium-sensitive electrode
(not included)
HI 7662 temperature probe
(not included)
4 x 1.5V AA size
200 hours of continuous use
0 to 50°C (32 to 122°F); 100% RH
196 x 80 x 60 mm (7.7 x 3.1 x 2.4")
450 g (15.8 lb.)
mannacl_r1.P65
6
6
FUNCTIONAL DESCRIPTION
HI 931101
1. Battery Cover
2. BNC Electrode connector
3. RCA Temperature Probe socket
4. LCD Display
5. ON/OFF key to turn the instrument on or off
6. CAL key to enter or exit calibration mode
7. MODE key, to select the scale (Na or pNa)
8. CFM key, to confirm calibration
9. BUF key, to select the calibration buffer value