User Guide

19
19
it with some of the buffer calibration
solution you are going to use first.
Lower the rubber sleeve to get the
junction to work properly
SINGLE POINT CALIBRATION (OFFSET)
Dip the FC 300B sodium
electrode into the 3.00 g/L
NaCl solution (HI 7083 or
HI 8083).
Note: the electrode should
be immersed ap-
proximately 2.5 cm
(1") into the solu-
tion.
Press the CAL key. The
"CAL" and "BUF" indica-
tors will be displayed with
the "BUF" and the "
"
symbols blinking.
As soon as the standard
solution is recognized, the
BUF symbol will stay still
while the "
" symbol
will continue to blink indi-
cating that the electrode
is stabilizing.
The value of the standard
solution being used will be
displayed on the primary
LCD.
As soon as the electrode
is stabilized, "CON" sym-
bol will appear on the dis-
play.
For best results, wait for a
couple of minutes before
CAL
CAL
3 g/L
NaCl
mannacl_r1.P65
20
20
pressing the CFM key. HI931100 and
HI931102 will then display "30.0 gm".
If the CAL key is pressed
at this point, the calibra-
tion process is ended pro-
viding only one point cali-
bration. For better accu-
racy however, it is recom-
mended that a two-point
calibration is performed.
TWO POINT CALIBRATION
Proceed as described in "Single Point Cali-
bration" but do not quit calibration by press-
ing the CAL key.
After the first point is con-
firmed, immerse the
FC 300B electrode into the
second buffer solution at
0.3 g/L NaCl (HI 7085/
HI8085) or 30 g/L NaCl
(HI7081/HI8081).
Note: the electrode should
be immersed ap-
proximately 2.5 cm
(1") into the solu-
tion.
CFM
CFM
gm
°C
NaCl
BUF
CAL
CAL
CAL
0.3 g/L or 30 g/L
NaCl