User Guide
14
pH
4 cm
(1½")
HI 7007
CAL
5. “4” can be changed to “10” (indicating pH
10.01) by pressing BUFF button. If you are
measuring acid samples (pH 7 or less) use
the pH 4.01 buffer (HI7004). For alkaline
solutions (pH 7 or greater) use pH 10.01
(HI7010).
6. Rinse the electrode with clean water or some
pH4.01 (or 10.01) solution. Dip the bottom 4
cm (1½") of the electrode in a beaker con-
taining the second buffer.
7. When “4” (or “10”) stops blinking, indicat-
ing that the electrode has stabilized, wait
30 seconds and then press CFM again to
confirm the second calibration point.
8. The buffer reference disappears from the
secondary display and the meter is cali-
brated and ready to use.
9. If the buffer is incorrect or the electrode is
defective, E2 will blink to alert the user.
Repeat the calibration procedure from the
beginning with fresh buffers. If you cannot
complete the calibration, check your elec-
trode by following the conditioning and maintenance procedure
on page 29 and repeat calibration. The pH electrode might have
to be replaced if calibration cannot be successfully performed.
PH CALIBRATION HI 92240
1. Rinse the electrode with some pH 7.01
(HI7007) solution or clean water. Dip the
bottom 4 cm (1½") of the electrode and the
temperature probe into a beaker contain-
ing pH7.01 solution.
2. Press the CAL key. The secondary (lower)
LCD will blink “7” to indicate that pH
7.01 is expected and the electrode is
conditioning.
BUFF
4 cm
(1½")
HI 7004
CFM










