User Guide

2726
Allow the solution to cool, and pour it
into the chamber marked "RH11.1%",
making sure that no residue remains on
the walls of the glass container.
Seal the chamber well when not in use,
as the LiCl solution is extremely hygro-
scopic and tends to capture the humidity
present in the air causing the solution to
expand in volume and overflow from the
container.
Pour approximately 12 cc of distilled wa-
ter into the other chamber marked "RH
75.4%".
Add all the content of the HI 7121
bottle containing NaCl while continuously
shaking the container to avoid the for-
mation of lumps. Seal this container well
when not in use.
The calibration kit needs 4 hours for proper
stabilization.
CALIBRATION PROCEDURE
Bring the calibration kit to a temperature of approximately 20°C
and keep it in an area with minimum amount of temperature
variations.
Remove the cap from the "RH 11.1%"
chamber containing the LiCl solution and
insert the probe paying attention that it
does not come in contact with the liquid.
Remove the adhesive sticker which covers
the calibration trimmer access holes.
Wait for the measurement to stabilize
(this takes approximately 4 hours).
L
i C
l
11
.1
%
N
a
C
l
7
5
.4
%
CALIBRATION KIT
L
i C
l
11
.
1
%
N
a
C
l
7
5
.4
%
CALIBRATION KIT
H
2
O
12cc
Na Cl
Li Cl
11.1%
Na Cl
75.4%
C
A
LIB
R
A
T
IO
N
K
IT
12
3
6
9
4 h
CALIBRATION KIT
L
i C
l
11
.1
%
N
a
C
l
7
5
.4
%
12
3
6
9
4 h
All Hanna thermohygrometers have been precalibrated at the factory.
Hanna Instruments uses state-of-the-art thermally-isolated humidity
chambers for this purpose.
It is generally recommended to have all thermohygrometers recalibrated
at least once a year.
For an accurate annual recalibration, contact your dealer or the
nearest Hanna
Customer Service Center.
You can also check the status of your thermohygrometer and perform
a quick RH recalibration (with an accuracy of ±5%) by using the
optional Hanna HI 7101
mini-calibration chamber.
The kit is composed of two thermally-isolated
chambers, each one equipped with a cap
and three bottles containing the appropriate
precalibrated saturated salts to produce a
known RH value.
PREPARING THE CALIBRATION SOLUTIONS
Pour approximately 26 cc of distilled wa-
ter into a glass container.
Immerse this container into a bath of ice
and water.
Slowly add the contents of a HI 7111
bottle containing LiCl into the glass con-
tainer while stirring gently.
When the salts have dissolved completely,
add all the content of the second bottle
of HI 7111.
CALIBRATIONCALIBRATION
CALIBRATIONCALIBRATION
CALIBRATION
L
i C
l
11
.1
%
N
a
C
l
7
5
.4
%
CALIBRATION KIT
H
2
O
26 cc
Li C
l
Li C
l