User Guide

1918
meter will print the measurements taken in
that moment, and will print with the indi-
cated printing interval thereafter until the
ON/OFF key is pressed.
The printout provides the following information:
a – A running log number
b – A running sample number in that particular log
c – Date (DD-MM-YY)
d – Printing interval in minutes
e – Time (HH.MM)
f – RH% value
g – Temperature value
When the meter is in recording mode "LOG" is
displayed on the secondary LCD with the RH%
value on the primary LCD.
Press the RANGE key to read the temperature
on the primary display.
If no keys are pressed after approx. 5 minutes, the meter goes in
standby mode to prolong the battery life.
Note: If PRINT is pressed while still in recording mode, a printout is
produced without affecting the running number.
Printing during recording
INTV
LOG
10-01-96
--01-- 0005M
0001 *11.05
65.6 RH%
25.0 ° C
<>
a
b
c
d
e
f
g
LOG
RANGE
10-01-96
0005M
-- 05 -- *11.51
65.2 RH%
25.0 ° C
PRINT
To view RH% on the primary display press RANGE.
To view temperature on the primary display when in RH% mode,
press RANGE again.
If “----" is displayed, it indicates that no probe
is connected (or that the probe cable is bro-
ken).
Note: the RH% value is displayed without any symbol on the
primary display. The temperature value is always displayed
together with the "°C" or "°F" symbol.
RANGE
°C
RANGE
PRINTING/RECORDINGPRINTING/RECORDING
PRINTING/RECORDINGPRINTING/RECORDING
PRINTING/RECORDING
WITH HWITH H
WITH HWITH H
WITH H
I I
I I
I
9161 & H9161 & H
9161 & H9161 & H
9161 & H
I I
I I
I
92619261
92619261
9261
To print the measured values press PRINT.
Each printout provides the following informa-
tion:
a – Running sample number
b – Date (DD-MM-YY)
c – Time (HH.MM)
d – RH% value
e – Temperature value
RECORDING AT AN INTERVAL
Set the appropriate logging interval (see pages
12-14). Press LOG to enter the recording mode.
The log number and the recording interval will
appear for a few seconds on the secondary
display to indicate the operational mode. The
PRINT
<>
10-01-96
0001 *10.59
65.6 RH%
25.0 ° C
a
b
c
d
e
LOG