Turn the instrument ON by pressing On/Off.
At start-up the display will show all
the used segments for a few seconds
(or while the button is held), followed
by the percentage indication of the
remaining battery life.
After a few seconds “Cond” message
appears, the “ ” and “WAIT” tags will
blink on the LCD if probe is connected
to inform the user that the probe is in
auto-conditioning (automatic polarization)
mode.
When this message disappears, the probe
is polarized and the instrument can be
calibrated.
The meter is now ready to operate.
If the probe is disconnected or broken, the
meter will display “----” and “NO Probe
tag blinking. In this situation only SETUP
menu is available.
The auto-off feature turns the instrument off
after a set period (default 20 min) with no
button pressed to save battery life. To set
another period or to disable this feature, see SETUP menu on page 17.
The auto-off backlight feature turns the backlight off after a set period
(default 1 min) with no buttons pressed. To set another period or to
disable this feature, see SETUP menu on page 17.
PROBE POLARIZATION
The probe is under polarization with a fixed voltage of approximately
800 mV.
Probe polarization is essential for stable measurements with the same
recurring degree of accuracy.