User Guide

19
19
PRINTING / RECORDING
WITH HI 9141
To print the measured values
press the PRINT key.
The printout provides the fol-
lowing information:
a - Running sample number
b - Date (DD-MM-YY)
c - Time (HH-MM)
d - D.O. value in ppm.
e - Temperature value
in °C.
RECORDING MODE (PROGRAMMED
PRINTOUTS)
Press the RECORD key to enter the record-
ing mode. The log number and the interval
time will appear for a few seconds on the
display to indicate the correct operational
mode.
The meter will print the measure-
ment taken in that moment, and
will print at the interval selected
thereafter until the RECORD key
is pressed.
Each printout provides the following informa-
tion:
a - A running log number
b - A running sample number (in that particu-
lar log)
PRINT
a
b
c
d
e
<>
14-01-96
0015 *00.50
8.1 ppm
20.4 ° C
LOG
<>
a
b
c
d
e
f
g
14-01-96
--02-- 0005M
0003 *15.50
8.5 ppm
25.4 º C
RECORD
RECORD
Untitled-22
20
20
c - Date (DD-MM-YY)
d - Printing interval indicator in minutes
e - Time (HH.MM)
f - D.O. value in ppm
g - Temperature value in °C.
When the meter is in record-
ing mode "LOG" is displayed
on the bottom left corner of
the LCD with the temperature
value on the primary display.
Notes:
It is recommended to use the external power
supply (HI710005 or HI710006) during
recording mode, especially when many
printouts are required.
Before proceeding with recording, make
sure there is enough paper for your
measurements. When the paper is fin-
ished the meter will not advise the opera-
tor and the printouts could be lost.
It is possible to insert a new paper roll
during recording session (see page 31).
If the PRINT key is pressed while
still in recording mode, a print-
out is produced without affecting
the running sample number.
Once in recording mode, the
printing interval cannot be
changed. Exit the recording
mode first (pressing the
RECORD key) before setting the new inter-
val (see page 13).
TO STOP RECORDING
In order to quit the recording mode, press the
RECORD key. This will generate a recording
exit status printout.
10-01-96
==07== 0005M
0016 *16.25
#1 20.0 °C
#2 57.3 °C
.
.
Log number
Total sample number
LOG EXIT
Date/Time
PRINT
°C
LOG
RECORD