User Guide
19
19
PRINTING / RECORDING
WITH HI 9141
To print the measured values
press the PRINT key.
The printout provides the fol-
lowing information:
a - Running sample number
b - Date (DD-MM-YY)
c - Time (HH-MM)
d - D.O. value in ppm.
e - Temperature value
in °C.
RECORDING MODE (PROGRAMMED
PRINTOUTS)
Press the RECORD key to enter the record-
ing mode. The log number and the interval
time will appear for a few seconds on the
display to indicate the correct operational
mode.
The meter will print the measure-
ment taken in that moment, and
will print at the interval selected
thereafter until the RECORD key
is pressed.
Each printout provides the following informa-
tion:
a - A running log number
b - A running sample number (in that particu-
lar log)
PRINT
a
b
c
d
e
<>
14-01-96
0015 *00.50
8.1 ppm
20.4 ° C
LOG
<>
a
b
c
d
e
f
g
14-01-96
--02-- 0005M
0003 *15.50
8.5 ppm
25.4 º C
RECORD
RECORD
Untitled-22
20
20
c - Date (DD-MM-YY)
d - Printing interval indicator in minutes
e - Time (HH.MM)
f - D.O. value in ppm
g - Temperature value in °C.
When the meter is in record-
ing mode "LOG" is displayed
on the bottom left corner of
the LCD with the temperature
value on the primary display.
Notes:
• It is recommended to use the external power
supply (HI710005 or HI710006) during
recording mode, especially when many
printouts are required.
• Before proceeding with recording, make
sure there is enough paper for your
measurements. When the paper is fin-
ished the meter will not advise the opera-
tor and the printouts could be lost.
• It is possible to insert a new paper roll
during recording session (see page 31).
• If the PRINT key is pressed while
still in recording mode, a print-
out is produced without affecting
the running sample number.
• Once in recording mode, the
printing interval cannot be
changed. Exit the recording
mode first (pressing the
RECORD key) before setting the new inter-
val (see page 13).
TO STOP RECORDING
In order to quit the recording mode, press the
RECORD key. This will generate a recording
exit status printout.
10-01-96
==07== 0005M
0016 *16.25
#1 20.0 °C
#2 57.3 °C
.
.
Log number
Total sample number
LOG EXIT
Date/Time
PRINT
°C
LOG
RECORD