Instruction Manual
2322
PRINT
CAL
LOG
LOG
The printer will print a complete set of data
and the "LOG" symbol is displayed.
Each printout provides the following informa-
tion:
a - Current log number
b - Current sample number (in that log)
c - Date (DD-MM-YY)
d - Printing interval indicator in minutes
e - Time (HH-MM)
f - DO value in ppm
g- Temperature
value in °C
It is always possible to switch from the log-
ging with printing function to the logging func-
tion. Press the CAL key and the "LOG" sym-
bol will start blinking to indicate that the data
are now stored only into memory.
Notes:
• It is recommended to use the external power
supply during logging with printing mode,
especially when many printouts are
required.
• Before proceeding with logging with print-
ing, make sure there is enough paper for
your measurements. When the paper is
finished the meter will not advise the op-
erator and the printouts could be lost.
Anyway, data will continue to be stored
into the meter memory.
• It is possible to insert a new paper roll
during logging session (see "Printer
Maintenance" section for details).
14-01-96
--03-- 0010M
0001 *19.10
8.02 ppm
22.1 °C
<>
a
b
c
d
e
f
g
• Once in logging mode, the
interval cannot be changed.
First exit the logging mode
by pressing the LOG key,
and then set the new in-
terval.
• If the PRINT key is
pressed while in logging
mode, a printout is pro-
duced without affecting the
running sample number.
TO STOP LOGGING
Press the LOG key, this will generate a log
exit status printout.
10-01-96
==07== 0005M
0016 *16.25
#1 20.0 °C.
Log number
Total sample number
LOG EXIT
Date/Time
14-01-96
- -09- - 0001M
0016 *20.00
<>