1918
a - Current log number
b - Current sample number (in that log)
c - Date (DD-MM-YY)
d - Printing interval indicator in minutes
e - Time (HH.MM)
f - D.O. value in ppm
g - Temperature value in °C.
When the meter is in record-
ing mode, "LOG" is displayed
on the LCD with the tempera-
ture value.
Notes:
It is recommended to use an external
power supply during recording mode, es-
pecially when many printouts are required.
Before proceeding with recording, make
sure there is enough paper. When the
paper is finished, the meter will not advise
the operator and printouts could be lost.
It is possible to insert a new paper roll
during recording session (see "Printer
Maintenance" section).
If PRINT is pressed while
in recording mode, a print-
out is produced without af-
fecting the running sample
number.
Once in recording mode, the printing inter-
val cannot be changed. Exit the recording
mode by pressing the
RECORD key, before set-
ting the new interval.
TO STOP RECORDING
In order to quit the recording mode, press the
RECORD key. This will generate a recording
exit status printout.
10-01-96
==07== 0005M
0016 *16.25
#1 20.0 °C
#2 57.3 °C
.
.
Log number
Total sample number
LOG EXIT
Date/Time
PRINT
°C
LOG
RECORD
To print the measured values, press
the PRINT key.
The printout provides the following informa-
tion:
a - Current sample number
b - Date (DD-MM-YY)
c - Time (HH-MM)
d- D.O. value in
ppm.
e- Temperature
value in °C.
RECORDING MODE (PROGRAMMED
PRINTOUTS)
Press the RECORD key to enter the record-
ing mode. The log number and the interval
time will be displayed for a few seconds to
indicate the correct operational mode.
The meter will print the measurement taken
in that moment, and will print at
the interval selected thereafter un-
til the RECORD key is pressed.
Each printout provides the following informa-
tion:
PRINT
LOG
RECORD
RECORD
PRINTING / RECORDING WITH HI 9141
a
b
c
d
e
<>
14-01-96
0015 *00.50
8.1 ppm
20.4 °C
<>
a
b
c
d
e
f
g
14-01-96
--02-- 0005M
0003 *15.50
8.5 ppm
25.4 ºC