User Guide

5352
Press .
Press 9
Press 9
Press 9. "0.999" will blink on the primary
LCD with the "CAL" symbol blinking
Press the ENTER key. "0.999" stays still
and the "CON" symbol starts blinking
Press the CON key to confirm and the
"CON" symbol will disappear
Note: To quit from the setting mode, press
the CAL key.
This program provides a fast way to perform
the calibration (without going through the "CAL"
procedure) based on the formula:
KNaCl = (CALBUF * (1 + 1.95% (Tp-25)) / Gt
CALBUF = calibrated buffer solution
Gt = measured solution conductivity at TP
Tp = temperature in °C
e.g. The buffer value is 100.0%, but the
meter shows 90.0%, so by changing
the KNaCl to 1.111, the meter will show
100.0% and will be completely cali-
brated.
* calibrated factor value range : 0.5 - 3.5
To revert to normal operation:
Press the PGM key
Press 0
Press the ENTER key.
This program provides a fast way to perform
the calibration (without going through the "CAL"
procedure) based on the formula:
KFERT = (CALBUF * (1 + 3.0% (Tp-25)) / Gt
where:
CALBUF = calibrated buffer solution
Gt = measured solution conductivity at Tp
Tp = temperature in °C
e.g. The buffer value is 1.00 g, but the
meter shows 0.9 g, so by changing the
KFert to 1.111, the meter will then
show 1.00 g and the meter is cali-
brated.
* calibrated factor value range : 0.01 - 1.50
To return to normal operation:
Press the PGM key
Press 0
Press the ENTER key.
PROGRAM # 9
To set or display the NaCl calibration factor
A. To display the NaCl calibration factor:
Press PGM
Press 9
Press ENTER and "P 9" and the current
NaCl calibration factor will respectively
appear on the TEMP. and on the primary
LCD
B. To set the NaCl calibration factor:
e.g. CAL factor 0.999
Press the CAL key and the current calibra-
tion factor will blink on the primary display.