PROGRAM # 8
To display or set Fert 1/2 calibration factor
A. To display the Fert 2 calibration factor
Press PGM
Press 8
Press ENTER. "P 8" and the current fertil-
izer calibration factor will respectively ap-
pear on the TEMP. and on the primary
LCD. The secondary display will show "2"
for Fertilizer #2
B. To change the fertilizer #:
Press the PGM key again
Press 8 again
Press the ENTER key. "P 8" and the
current fertilizer calibration factor will re-
spectively appear on the TEMP. and on
the primary LCD. The secondary display
will show "1" for Fertilizer #1
C. To set the Fert 1 calibration factor:
e.g. CAL factor 0.999
Press CAL and the current setting will
blink on the primary LCD
Press .
Press 9
Press 9
Press 9 and "0.999" will blink on the pri-
mary LCD
Press the ENTER key. "0.999" stays still
and the "CON" symbol will blink
Press the CON key to confirm and the
"CON" symbol will disappear
Note: To quit from the setting mode, press
the CAL key.
B. Setting the conductivity calibration factor
e.g. CAL factor is 0.999
Press the CAL key and the current calibra-
tion factor will blink on the primary LCD
Press .
Press 9
Press 9
Press 9. "0.999" and "CAL" symbol will
blink on the large LCD
Press ENTER. "0.999" will be fixed on
large LCD with the "CON" symbol will be
blinking
Press the CON key
Note: To quit from the setting mode, press
the CAL key.
This program provides a fast way to perform
the calibration (without going through the cali-
bration procedure) based on the formula:
KCond = (CALBUF * (1 + TC (TP - 25 )) / Gt
where:
CALBUF = calibrated buffer solution
Gt = measured solution conductivity at Tp
TC = temperature coefficient
TP = sample temperature in °C
e.g. The buffer value is 1000 µS/cm, but the
meter shows 900 µS/cm, so by chang-
ing the KCond to 1.111, the meter will
then show 1000 µS/cm and the meter
is calibrated.
* calibration factor value range: 0.850 - 1.150
To return to normal operation:
Press the PGM key
Press 0
Press the ENTER key.