User Guide

4948
PROGRAM # 2
Setting of RS232 ESC code
e.g. ESC code is 20 (decimal)
Press PGM
Press 2
Press ENTER. "P 2" on the TEMP. display
and the current setting will blink on the
primary LCD
Press 2
Press 0. "20" will blink on the primary LCD
Press ENTER. "20" will be fixed on the primary
LCD
To return to normal operation:
Press the PGM key
Press 0
Press the ENTER key.
PROGRAM # 3
Resetting RS232 code and Baud Rate to the
default value
Press PGM
Press 3
Press ENTER. "P 0" will blink on the
primary LCD
To return to normal operation press the EN-
TER key.
PROGRAM # 7
To display or set conductivity calibration factor
A. To display the conductivity calibration factor
Press PGM
Press 7
Press ENTER. "P 7" and the current con-
ductivity calibration factor will appear re-
spectively on the TEMP. and on the pri-
mary LCD
Press 1 and "P1"
will blink
Press ENTER
and "P 1" will be
displayed on the
TEMP. LCD
To stop any of the programs or to quit to
revert to normal operation:
Press the PGM key
Press 0
Press the ENTER key.
PROGRAM # 1
Setting of RS232 Baud Rate
e.g. Baud Rate 300
Press PGM
Press 1
Press ENTER. "P 1" on the TEMP. dis-
play and the current setting will blink on
the primary LCD
Press 3
Press 0
Press 0. "300" will blink on the primary LCD
Press ENTER. "300" will be fixed on the pri-
mary LCD
To return to normal operation:
Press the PGM key
Press 0
Press the ENTER key.
FTR
1
ENTER
PGM
Clear
ATC
MTC
0
ENTER