User Guide

3332
Notes:
If the temperature probe is not con-
nected to the meter, once the selected
buffer value is confirmed, "25°C" or
"77°F" will start blinking on the TEMP.
display.
Enter the sample tempera-
ture value using the numeri-
cal keys and press the EN-
TER key to confirm the tem-
perature setting.
Now the calibration procedure is com-
plete.
To quit from the calibration
mode, press the CAL key.
To clear the numeric data
entry, press the CLEAR key.
PROCEDURE WITHOUT ATC
(MTC)
If you need to calibrate without ATC,
press the ATC/MTC key and the deci-
mal point will blink on the TEMP. LCD
(both in the Celsius and in the Fahren-
heit scale).
During calibration once the selected
buffer value is confirmed, "25°C" or
"77.0°F" will start blinking on the TEMP.
display to prompt the selection of the
calibration temperature value.
ENTER
CAL
PGM
Clear
ATC
MTC
0
the liquid. Immerse the
HI 7669/2W temperature
probe as close as pos-
sible to the conductivity
probe. Wait for 2 or 3 min-
utes for the thermal equi-
librium.
If the buffer is within
39 and 49 mS, the
"
BUF
" symbol stays
still.
If the " " starts
blinking, wait for ap-
proximately 30 sec-
onds for the reading
to stabilize.
Once the reading is
stable the " "
symbol goes off and
CON prompts the
user to confirm the
reading.
Press the CON key to
confirm the calibration.
The instrument is now calibrated and returns
to normal operation.
CON
NaCl
%
BUF
CON
NaCl
CAL
%
BUF
NaCl
CAL
%
BUF