User Guide

3130
INITIAL PREPARATION
Rinse the probe thoroughly in distilled water.
This is to minimize contamination of the cali-
bration solution and secure higher accuracy.
Pour a small quantity of
HI7037 Standard solution into
a plastic beaker or any other
solution to be used as 100%
reference.
When possible, use plastic
beakers to minimize any EMC
interferences.
PROCEDURE WITH ATC
Press the NaCl key to enter the sea water
percentage mode, and the "NaCl" symbol
will appear on the large display.
Press CAL and the LCD will display
"100.0%".
Press the ENTER key and im-
merse the HI7686 probe into
the calibration solution submerg-
ing the holes of the sleeve into
NaCl CALIBRATION
NaCl
%
NaCl
9
ENTER
CAL
NaCl
CAL
%
(both in the Celsius and in the Fahren-
heit scale).
During calibration once
the selected buffer value
is confirmed, "25°C" or
"77.0°F" will start blink-
ing on the TEMP. dis-
play to prompt the se-
lection of the calibration
temperature value.
Take the temperature of
the solution with a
Checktemp or a glass thermometer.
Enter the correct buffer temperature
value by using the numeric keys and
press ENTER to confirm the tempera-
ture setting.
Continue the procedure as for calibra-
tion with ATC, but without temperature
probe connection.
ENTER