User Guide

2928
If the HI7669/2W temperature probe is
not connected to the meter, once the
selected buffer value is confirmed,
"25°C" or "77.0°F" will start blinking on
the TEMP. display to prompt the se-
lection of the calibration temperature
value.
Take the temperature
of the solution with a
Checktemp or a glass
thermometer.
Enter the correct buffer
temperature value by
using the numeric keys
and press ENTER to
confirm the temperature
setting.
Now the calibration procedure is com-
plete.
To quit the calibration mode,
press the CAL key.
To clear the numeric data
entry, press CLEAR key.
The instrument should be calibrated
weekly or every time the probe has
been changed.
PROCEDURE WITHOUT ATC (MTC)
If you need to calibrate without ATC,
press the ATC/MTC key and the deci-
mal point will blink on the TEMP. LCD
ENTER
CAL
PGM
Clear
CONDUCTIVITY
CAL
%TC
S
µ
BUF
CONDUCTIVITY
%TC
S
µ
BUF
CON
CON
CONDUCTIVITY
CAL
%TC
S
µ
BUF
CAL
The "CAL" and "
BUF
"
prompts will appear
on the large display
and the " " sym-
bol will start blink-
ing. Wait for ap-
proximately 30 sec-
onds for the reading
to stabilize.
When the value is
stable, " " and
"CAL" symbols will
disappear and the
"CON" symbol will
start blinking to
prompt the user to
confirm the reading.
Press the CON key
to confirm the cali-
bration.
The conductivity calibration procedure is now
complete.
Notes:
If the buffer value
differs by more
than ±15% from
the ideal value,
the "
BUF
" symbol
will blink on the
display.
Press the CAL
key to exit the
calibration mode
and repeat the
procedure using
the correct buffer.